Chapter 4 Command Reference
Introduction
Programming Guide
4-21
Stats
:CALCulate3:AVERage:COUNt:CURRent?
Queries the current count (that is, the number of data values collected for
statistical computation).
•
Numeric data transferred as ASCII bytes in <NR1> format.
•
Range is 0 to 1,000,000.
•
If :CALC3:AVER[:STATe] is OFF, error -221 is generated.
•
If the current measurement is Totalize or Voltage Peaks, 0 is returned and
error -221 is generated.
•
Query only.
•
No statistics results exist until the :CALC3:AVER:COUN:CURR? is equal to
the specified :CALC3:AVER:COUN.
Stats
:CALCulate3:AVERage[:STATe] <Boolean>
Sets or queries the statistics post-processing enable .
•
Single ASCII-encoded byte, 0 or 1.
•
A value of 0 indicates OFF; a value of 1 indicates ON.
•
*RST: OFF
•
When this enable is ON, and :TRIG:COUN:AUTO is ON, and
[:SENS]:FUNC[:ON] is not Totalize or Voltage Peak function,
then :INIT[:IMM] initiates a complete block of measurements.
See :TRIG:COUN:AUTO in this chapter for specifics.
•
When this enable is OFF, :INIT[:IMM] always initiates a single measurement.
Related
Front-Panel
keys
Query Response
Comments
Related
Front-Panel
keys
Query Response
Comments
Summary of Contents for 53131A
Page 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Page 2: ......
Page 13: ...1 1 Before You Start ...
Page 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Page 25: ...2 2 Command Summary A Quick Reference ...
Page 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Page 69: ...3 3 Programming Your Universal Counter for Remote Operation ...
Page 167: ...4 4 Command Reference A Dictionary ...
Page 310: ...Chapter 4 Command Reference WAI Wait to Continue Command 4 144 Programming Guide ...
Page 311: ...5 5 Errors ...
Page 329: ......
Page 330: ...H Manual Part Number 53131 90044 Printed in U S A NOVEMBER 1996 ...