Chapter 4 Command Reference
Introduction
Programming Guide
4-49
:INPut[1|2]:FILTer[:LPASs]:FREQuency?
Queries the cutoff frequency of the low-pass filter.
•
Numeric data transferred as ASCII bytes in <NR3> format with six
significant digits.
•
A value of 100E+3 is returned.
Units are Hertz.
:INPut[1|2]:IMPedance <numeric_value> [OHM]
Sets or queries the input impedance (50
Ω
or 1M
Ω
).
50 or 1E6
Numeric data transferred as ASCII bytes in <NR3> format with six significant
digits.
•
*RST: 1E6 OHM
•
Units are Ohms.
50
Ω
/1M
Ω
Query Response
Comments
<numeric_value>
Range
Query Response
Comments
Related
Front-Panel
Keys