Chapter 2 Command Summary
HP 53131A/132A Command Summary
2-22
Programming Guide
Table 2-1. IEEE 488.2 Common Commands
Mnemonic
Command Name
Function
*CAL?
*CLS
*DDT <arbitrary block>
*DMC <string>,
<arbitrary block>
*EMC <NRf>
*EMC?
*ESE <NRf>
*ESE?
*ESR?
*GMC? <string>
*IDN?
*LMC?
*OPC
*OPC?
Calibration
Clear Status
Define Device Trigger Command
Define Macro Command
Enable Macro Command
Enable Macro Query
Standard Event Status Enable
Standard Event Status Enable
Query
Event Status Register Query
Get Macro Contents Query
Identification Query
Learn Macro Query
Operation Complete
Operation Complete Query
Causes the Counter to perform an internal interpolator self-
calibration and returns a response that indicates whether or
not the instrument completed the self-calibration without
error.
Clears Status data structures (Event Registers and Error
Queue).
Defines either INIT, FETC?, READ?, or nothing to be
executed when the Counter receives a GET or *TRG
command.
Assigns a sequence of zero or more commands/queries to a
macro label.
No query form.
Enables and disables expansion of macros.
Non-zero value enables; zero value disables.
Queries whether macros are enabled.
Sets the Standard Event Status Enable Register.
Queries the Standard Event Status Enable Register.
Queries the Standard Event Status Register.
Queries the current definition of a currently defined macro
label.
Queries the Counter identification.
Queries the currently defined macro labels.
Causes Counter to set the operation complete bit in the
Standard Event Status Register when all pending
operations (see Note) are finished.
Places an ASCII “1” in the Output Queue when all
pending operations (see Note) are completed.
Note: Pending operations include measurements in progress.
Summary of Contents for 53131A
Page 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Page 2: ......
Page 13: ...1 1 Before You Start ...
Page 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Page 25: ...2 2 Command Summary A Quick Reference ...
Page 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Page 69: ...3 3 Programming Your Universal Counter for Remote Operation ...
Page 167: ...4 4 Command Reference A Dictionary ...
Page 310: ...Chapter 4 Command Reference WAI Wait to Continue Command 4 144 Programming Guide ...
Page 311: ...5 5 Errors ...
Page 329: ......
Page 330: ...H Manual Part Number 53131 90044 Printed in U S A NOVEMBER 1996 ...