Chapter 4 Command Reference
:INITiate Subsystem
4-46
Programming Guide
•
When a single measurement is in progress (:INIT:CONT is OFF):
– Error -213 (Init ignored) is generated and the state of
INIT:CONT is unaffected by :INIT:CONT ON.
– Error -210 (Trigger error) is generated by INIT:CONT OFF.
•
Note that the Counter powers up with :INIT:CONT set to ON,
but *RST sets :INIT:CONT to OFF.
Run
:INITiate[:IMMediate]
This event command causes the instrument to initiate either a single measurement
or a block of measurements.
When
:TRIG:COUN:AUTO is OFF, or
:CALC3:AVER[:STAT] is OFF, or
[:SENS]:FUNC[:ON] is Totalize function, or
[:SENS]:FUNC[:ON] is any Voltage Peaks function,
then :INIT[:IMM] initiates a single measurement.
When
:TRIG:COUN:AUTO is ON, and
:CALC3:AVER[:STAT] is ON, and
[:SENS]:FUNC[:ON] is not Totalize or Voltage Peak function,
then :INIT[:IMM] initiates a complete block of measurements.
See :TRIG:COUN:AUTO for specifics.
Related
Front-Panel
Keys
Summary of Contents for 53131A
Page 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Page 2: ......
Page 13: ...1 1 Before You Start ...
Page 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Page 25: ...2 2 Command Summary A Quick Reference ...
Page 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Page 69: ...3 3 Programming Your Universal Counter for Remote Operation ...
Page 167: ...4 4 Command Reference A Dictionary ...
Page 310: ...Chapter 4 Command Reference WAI Wait to Continue Command 4 144 Programming Guide ...
Page 311: ...5 5 Errors ...
Page 329: ......
Page 330: ...H Manual Part Number 53131 90044 Printed in U S A NOVEMBER 1996 ...