Chapter 3 Programming Your Universal Counter for Remote Operation
Status Reporting
3-32
Programming Guide
Note that this is the only bit in the Operation Status Register which is not
representing a
condition
. Therefore, the transition filters have no effect on
this bit.
The Counter does not monitor the condition indicating whether the last
measurement was in or out of limit. Hence, the In Limit Event bit does
NOT represent the transition from an “out of limit measurement” to “in
limit measurement.”
•
Bits 11–15
are not used.
Questionable Data/Signal Status Register Group
The Questionable Data/Signal Status Register Group monitors
SCPI-defined conditions.
Table 3-8 lists the Questionable Data/Signal Status Register bits and briefly
describes each bit.
Table 3-8. Questionable Data/Signal Status Register
BIT
WEIGHT
DESCRIPTION
0
Not used
1
Not used
2
4
Time (Time Interval, Rise Time, Fall Time, Pulse Width,
Duty Cycle, and Period)
3
Not used
4
Not used
5
32
Frequency (only Frequency; not Frequency Ratio)
6
64
Phase
7
Not used
8
256
Calibration Error
9
Not used
10
1024
Out of Limit Event
11–13
Not used
14
16384
Command Warning
15
Not used since some controllers may have difficulty
reading a 16-bit unsigned integer. The value of this bit
shall always be 0.
Summary of Contents for 53131A
Page 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Page 2: ......
Page 13: ...1 1 Before You Start ...
Page 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Page 25: ...2 2 Command Summary A Quick Reference ...
Page 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Page 69: ...3 3 Programming Your Universal Counter for Remote Operation ...
Page 167: ...4 4 Command Reference A Dictionary ...
Page 310: ...Chapter 4 Command Reference WAI Wait to Continue Command 4 144 Programming Guide ...
Page 311: ...5 5 Errors ...
Page 329: ......
Page 330: ...H Manual Part Number 53131 90044 Printed in U S A NOVEMBER 1996 ...