Chapter 4 Command Reference
Introduction
Programming Guide
4-11
:CALCulate2 Subsystem
This subsystem performs post-acquisition limit testing and data transfer.
Not until :CALCulate2:LIMit:STATe is set to ON will any of the :CALCulate2
settings be used.
:CALCulate2:FEED “[:]CALCulate[1]”
Sets or queries the data flow to be fed into the CALCulate2 block.
The string “CALC” is returned.
*RST: “CALCulate[1]”
:CALCulate2:IMMediate
This command is an event that causes the Counter to recalculate existing data
without re-acquiring data. (This recalculation also happens automatically when
any change is made to the :CALCulate[1|2] subsystems while
:CALC2:IMM:AUTO is ON.)
The only limit result that can be truly post-processed is :CALC2:LIM:FAIL?. The
limit counts (:CALC2:LIM:FCO and :CALC2:LIM:PCO) reflect measurements
that were limit-tested at time of data acquisition.
This command will not affect:
:CALC2:LIM:FCO
:CALC2:LIM:PCO
:CALC3: ...
This command causes post-processing to occur in the :CALCulate[1] subsystem,
as well as the :CALCulate2 subsystem.
NOTE
Query Response
Comments
Comments
Summary of Contents for 53131A
Page 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Page 2: ......
Page 13: ...1 1 Before You Start ...
Page 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Page 25: ...2 2 Command Summary A Quick Reference ...
Page 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Page 69: ...3 3 Programming Your Universal Counter for Remote Operation ...
Page 167: ...4 4 Command Reference A Dictionary ...
Page 310: ...Chapter 4 Command Reference WAI Wait to Continue Command 4 144 Programming Guide ...
Page 311: ...5 5 Errors ...
Page 329: ......
Page 330: ...H Manual Part Number 53131 90044 Printed in U S A NOVEMBER 1996 ...