Chapter 4 Command Reference
Introduction
Programming Guide
4-13
:CALCulate2:LIMit:CLEar:AUTO <Boolean>
Sets or queries if the limit test results are to be cleared with each
:INITiate[:IMMediate] and :INITiate:CONTinuous ON operation.
•
Single ASCII-encoded byte, 0 or 1.
•
A value of 0 indicates OFF; a value of 1 indicates ON.
•
*RST: ON
•
When AUTO is ON, the Counter will perform the following whenever
:INIT[:IMM] or :INIT:CONT ON is executed:
– Invalidate the limit data.
– Clear :CALC2:LIM:FAIL, :CALC2:LIM:FCOunt, and
:CALC2:LIM:PCOunt information.
– Turn off the front-panel display’s
Limit
annunciator.
– Set the Limit-Detect output of the RS-232 connector to the
in-limit voltage level.
•
When AUTO is OFF, the only way to clear the limit-test results is to send
:CALC2:LIM:CLE[:IMM].
:CALCulate2:LIMit:CLEar[:IMMediate]
This command is an event that causes the Counter to
•
immediately invalidate the limit data,
•
clear the information in :CALC2:LIM:FAIL, :CALC2:LIM:FCOunt, and
:CALC2:LIM:PCOunt,
•
turn off the front-panel display’s Limit annunciator, and
•
set the Limit-Detect output to its in-limit voltage level.
If :CALC2:LIM:STAT is OFF, error -221 is generated.
Query Response
Comments
Comments
Summary of Contents for 53131A
Page 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Page 2: ......
Page 13: ...1 1 Before You Start ...
Page 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Page 25: ...2 2 Command Summary A Quick Reference ...
Page 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Page 69: ...3 3 Programming Your Universal Counter for Remote Operation ...
Page 167: ...4 4 Command Reference A Dictionary ...
Page 310: ...Chapter 4 Command Reference WAI Wait to Continue Command 4 144 Programming Guide ...
Page 311: ...5 5 Errors ...
Page 329: ......
Page 330: ...H Manual Part Number 53131 90044 Printed in U S A NOVEMBER 1996 ...