Chapter 4 Command Reference
Introduction
Programming Guide
4-59
Table 4-1. The <function>, associated <parameters> and
<source_list> for the Measure Instruction Commands
<function> *
<parameters>
[,<source_list>]**
[:VOLTage]:DCYCle
or
[:VOLTage]:PDUTycycle
[<reference>]
[(@1)]
[:VOLTage]:FALL:TIME
or
[:VOLTage]:FTIMe
[<lower_reference> [,upper_reference>]]
[(@1) ]
[:VOLTage]:FREQuency
[<expected_value>[,<resolution>]]
[(@1) | (@2) | (@3)]
[:VOLTage]:FREQuency:RATio
[<expected_value> [,<resolution>]]
[(@1), (@2 | @3)] |
[(@2 | @3), (@1)]
[:VOLTage]:MAXimum
[(@1) | (@2)]
[:VOLTage]:MINimum
[(@1) | (@2)]
[:VOLTage]:NWIDth
[<reference>]
[(@1) ]
[:VOLTage]:PERiod
[<expected_value>[,<resolution>]]
[(@1) | (@2) | (@3)]
[:VOLTage]:PHASe
[(@1), (@2)]
[:VOLTage]:PTPeak
[(@1) | (@2)]
[:VOLTage]:PWIDth
[<reference>]
[(@1) ]
[:VOLTage]:RISE:TIME
or
[:VOLTage]:RTIMe
[<lower_reference> [,upper_reference>]]
[(@1) ]
[:VOLTage]:TINTerval
[(@1), (@2)]
[:VOLTage]:TOTalize:CONTinuous***
[(@1)]
[:VOLTage]:TOTalize:TIMed
[<gate_time>]
[(@1)]
_______________________
*
The only functions which can be derived (using FETC? or READ?) from the stored data are period
to/from frequency, maximum to/from minimum, maximum to/from peak-to-peak
and minimum to/from peak-to-peak. All other functions require an acquisition of a new type.
**
<source_list> has the same syntax as SCPI <channel _list> syntax. For example, a single-channel
function (e.g., frequency, period, etc.) would use (@1) to specify channel 1, where as a two-channel
function (e.g., time interval, phase, and frequency ratio) would use (@1), (@2) to specify a
measurement between channel 1 and channel 2.
***
This <function> is only allowed with :CONFigure.
Summary of Contents for 53131A
Page 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Page 2: ......
Page 13: ...1 1 Before You Start ...
Page 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Page 25: ...2 2 Command Summary A Quick Reference ...
Page 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Page 69: ...3 3 Programming Your Universal Counter for Remote Operation ...