Chapter 4 Command Reference
Introduction
Programming Guide
4-25
-9.999912 to -1.0000000000E-13, 0.0000000000, +1.0000000000E-
13 to +9.999912.
11 digits
Numeric data transferred as ASCII bytes in <NR3> format with eleven significant
digits.
•
*RST: 0.0000000000
•
This command couples :CALC2:LIM:UPP to the same value.
•
Updating the upper limit value causes the limit counts (:CALC2:LIM:FCO,
:CALC2:LIM:PCO) to be cleared.
•
The front panel menu item is not always able to display all of the significant
digits of this value. When this is the case, the displayed value is different from
the actual value in that the displayed value has been rounded. However, using
the front panel
Enter
key, while this value is in the 11-digit display, will
update the actual value to the displayed (rounded) value.
:CALCulate3:PATH?
Queries the order in which CALCulate3 sub-blocks are to be processed.
For the Counter, this sequence is fixed to be LFILter followed by AVERage.
A sequence of ASCII-encoded bytes: LFIL, AVER
Query only.
<numeric_value>
Range
<numeric_value>
Resolution
Query Response
Comments
Query Response
Comments
Summary of Contents for 53131A
Page 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Page 2: ......
Page 13: ...1 1 Before You Start ...
Page 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Page 25: ...2 2 Command Summary A Quick Reference ...
Page 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Page 69: ...3 3 Programming Your Universal Counter for Remote Operation ...
Page 167: ...4 4 Command Reference A Dictionary ...
Page 310: ...Chapter 4 Command Reference WAI Wait to Continue Command 4 144 Programming Guide ...
Page 311: ...5 5 Errors ...
Page 329: ......
Page 330: ...H Manual Part Number 53131 90044 Printed in U S A NOVEMBER 1996 ...