Chapter 4 Command Reference
*OPT?
(Option Identification Query)
4-134
Programming Guide
*OPT?
(Option Identification Query)
*OPT?
Queries the instrument to identify any installed options.
The following options can be installed in the instrument:
•
Option 001, Medium Stability Oven Timebase
•
Option 010, High Stability Oven Timebase
•
Option 012, Ultra High Stability Oven Timebase
•
Option 030, 3.0 GHz RF Input Channel (Channel 3)
•
Option 050, 5.0 GHz RF Input Channel (Channel 3)
•
A sequence of ASCII-encoded bytes, indicating <timebase option>, <3.0 GHz
RF Input option> terminated with a new line and EOI.
– The <timebase option> is 001, 010, or 0.
Note that the counter responds with 010 for either the high stability or ultra
high stability oven timebase.
– The <3.0 GHz RF Input option> is 030 or 0.
Note that the counter responds with 030 for either the 3.0 GHz or 5.0 GHz RF
Input option.
– A missing option is identified by an ASCII 0 (zero).
For example, if only the medium stability timebase option is detected, the response
would be: 001,0.
This query should be the last query in a terminated program message; otherwise,
error -440 is generated.
Query Response
Comments
Summary of Contents for 53131A
Page 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Page 2: ......
Page 13: ...1 1 Before You Start ...
Page 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Page 25: ...2 2 Command Summary A Quick Reference ...
Page 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Page 69: ...3 3 Programming Your Universal Counter for Remote Operation ...
Page 167: ...4 4 Command Reference A Dictionary ...
Page 310: ...Chapter 4 Command Reference WAI Wait to Continue Command 4 144 Programming Guide ...
Page 311: ...5 5 Errors ...
Page 329: ......
Page 330: ...H Manual Part Number 53131 90044 Printed in U S A NOVEMBER 1996 ...