Chapter 4 Command Reference
Introduction
Programming Guide
4-69
Descriptions of the Measurement Functions— <function> (Cont.)
:MEASure[:SCALar][:VOLTage]:PERiod?
(Cont.)
Ch3<expected_value>
range:
0.33 ns to 10.0 ns
resolution:
<expected_value> should be within 10% of input period
for optimum arming configuration
default:
2 ns
Ch3 <resolution>
description:
value indicates decade corresponding to least significant
digit of the result
range:
1E-24 to 1E-11 sec
value which indicates 3 to 15 digits of resolution for the
specified <expected value>
resolution:
<resolution> should use a mantissa of 1.0 and be an even
power of 10
default:
value which indicates 4 digits of resolution for the
specified <expected_value>
<source_list>
range:
(@1) | (@2) | (@3)
default:
(@1)