Chapter 4 Command Reference
Introduction
Programming Guide
4-89
•
*RST: ON
•
This value is unaffected by save/recall.
•
While the Counter is configured to ON, representative CW signal(s) must be
present at the measurements input(s).
•
The ON setting causes the Counter to disregard any previously set “expected
frequency” ([:SENS]:FREQ:EXP[1|2|3]).
•
The only mechanism for disabling the above described
pre-measurement step is to specify an expected frequency with
[:SENS]:FREQ:EXP[1|2|3].
[:SENSe]:FUNCtion[:ON] <sensor_function>
Sets or queries the <sensor_function> to be sensed by the Counter.
The <sensor_function> strings are :
“[:][XNONe:]DCYCle [1]”
or
“[:][XNONe:]PDUTycycle [1]”
“[:][XNONe:]FALL:TIME [1] ”
or
“[:][XNONe:]FTIMe [1]”
“[:][XNONe:]FREQuency [1|2|3] ”
“[:][XNONe:]FREQuency:RATio [ 1,2 | 1,3 | 2,1 | 3,1 ] ”
“[:][XNONe:]NWIDth [1]”
“[:][XNONe:]PERiod [1|2|3] ”
“[:][XNONe:]PHASe [1,2]”
“[:][XNONe:]PWIDth [1]”
“[:][XNONe:]TINTerval [1,2]”
“[:][XNONe:]TOTalize [1]”
“[:][XNONe:]RISE:TIME [1]”
or
“[:][XNONe:]RTIMe [1]”
“[:][XNONe:]VOLTage:MAXimum [1|2] ”
“[:][XNONe:]VOLTage:MINimum [1|2] ”
“[:][XNONe:]VOLTage:PTPeak [1|2] ”
Comments
Summary of Contents for 53131A
Page 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Page 2: ......
Page 13: ...1 1 Before You Start ...
Page 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Page 25: ...2 2 Command Summary A Quick Reference ...
Page 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Page 69: ...3 3 Programming Your Universal Counter for Remote Operation ...
Page 167: ...4 4 Command Reference A Dictionary ...
Page 310: ...Chapter 4 Command Reference WAI Wait to Continue Command 4 144 Programming Guide ...
Page 311: ...5 5 Errors ...
Page 329: ......
Page 330: ...H Manual Part Number 53131 90044 Printed in U S A NOVEMBER 1996 ...