Chapter 1 Before You Start ...
Programming Guide Contents
Programming Guide
1-5
:CONFigure:TOTalize:TIMed
:CONFigure:TOTalize:CONTinuous
:MEASure:TOTalize:TIMed?
If your Counter contains firmware revision s 3402 and below, the Totalize
Measurement Instruction commands (shown above) are not available to disable
auto-trigger.
In the firmware revisions 3402 and below, these commands enabled auto-trigger at
the 50% level.
HP 53132A Time Interval De lay Arming
HP 53131A and HP 53132A Counters with a serial number prefix below 3646 are
identical in their TI arming modes. Both only offer Time Interval Delay, where the
STOP trigger of a time interval measurement can be delayed by a user -specified
time.