Chapter 4 Command Reference
Introduction
Programming Guide
4-63
Descriptions of the Measurement Functions— <function> (Cont.)
:MEASure[:SCALar][:VOLTage]:FALL:TIME?
(Cont.)
<source_list>
range:
(@1)
default:
(@1)
:MEASure[:SCALar][:VOLTage]:FREQuency?
[<expected_value>[,<resolution>]][, (@1)|(@2)|(@3)]
Measures Frequency.
The measurement arming mode is set to “digits.” The Counter uses the
<expected_value> and <resolution> parameters to configure the number of digits
of resolution arming setting.
The Channel 1 or Channel 2 trigger settings are coupled so that the measurement
channel has auto-trigger enabled at 50% with a positive slope.
Ch1, Ch2
<expected_value>
range:
.100 Hz to 225 MHz
resolution:
<expected value> should be within 10% of input
frequency for optimum arming configuration
default:
10 MHz
Ch1, Ch2 <resolution>
description:
value indicates decade corresponding to least significant
digit of the result
range:
1E-16 to 1E6 Hz
value which indicates 3 to 15 digits of resolution for the
specified <expected value>
resolution:
<resolution> should use a mantissa of 1.0 and be an
even power of 10
Summary of Contents for 53131A
Page 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Page 2: ......
Page 13: ...1 1 Before You Start ...
Page 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Page 25: ...2 2 Command Summary A Quick Reference ...
Page 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Page 69: ...3 3 Programming Your Universal Counter for Remote Operation ...
Page 167: ...4 4 Command Reference A Dictionary ...
Page 310: ...Chapter 4 Command Reference WAI Wait to Continue Command 4 144 Programming Guide ...
Page 311: ...5 5 Errors ...
Page 329: ......
Page 330: ...H Manual Part Number 53131 90044 Printed in U S A NOVEMBER 1996 ...