Event
ID
Description
Possible Cause
Resolution
210
Optics drive unusually low
Optics LED not properly calibrated
Perform Replace CC or Adjust Optics operation to recalibrate
Optics board defective
Replace the Optics Block. Contact Honeywell service
Analyzer CPU defective
Replace Analyzer, Contact Honeywell service
211
Optics drive unusually high
Optics LED not properly calibrated
Perform Replace CC or Adjust Optics operation to recalibrate
LED degradated
Replace the optics block
Optics board defective
Replace the Optics Block. Contact Honeywell service
Analyzer CPU defective
Replace Analyzer, Contact Honeywell service
212
Excess optics signal
Optics LED not properly calibrated
Perform Replace CC or Adjust Optics operation to recalibrate
Optics board defective
Replace the Optics Block. Contact Honeywell service
Analyzer CPU defective
Replace Analyzer, Contact Honeywell service
213
Double Pump failure
See causes for Fault 109
See recoveries for Fault 109
Circuit breaker tripped
Contact Honeywell service
214
Flow gone for the specific point
Improper flow setup
Adjust flow to 200cc
Tape tracking problem
Reload Chemcassette
Condensation in system
Purge internal lines
Clogged micro tube
Replace tube.
Note: Calibration required, contact Service.
Gate not fully closing
Gate adjustment loose, pivot binding
Optics block loose
Tighten optics block fasteners
Flow adjustment is unstable during adjusting
Replace needle valves
215
Failed to stop Analyzers monitoring
mode
Queen Analyzer (first in the rack) failed to communicate with other
Analyzers
Replace first Analyzer in the rack
Analyzer hardware failure
Replace Analyzer, Contact Honeywell service
216
Point Pressure Out Of Range
Miscalibrated sensor board or defective transducer
Replace Analyzer
217
Supply Pressure Out Of Range
Miscalibrated sensor board or defective transducer
Replace Analyzer
218
Inadequate Analyzer Vacuum
Improper system pressure adjustment
Adjust system pressure to 11 inHg
219
Optics SW corrupted
Hardware failure
Replace the Optics Block
220
Optics Internal SW errors
Cable issue
Confirm cable connections
Hardware failure
Replace the Optics Block. Contact Honeywell service
221
LED adjustment failed due to
insufficient optical signal
Optic block dirty
Clean optics.
Contact Honeywell Analytics Service.
Vertex Edge System
134
User Manual
Summary of Contents for Vertex Edge
Page 1: ...Vertex Edge 72 POINT CONTINUOUS MONITOR ...
Page 2: ......
Page 6: ......
Page 19: ...Chemcassette 1 Chemcassette directional flow Vertex Edge System 18 User Manual ...
Page 29: ...Installation Drawings Vertex Edge System 28 User Manual ...
Page 46: ......
Page 59: ...Vertex Edge System 58 User Manual ...
Page 63: ...Vertex Edge System 62 User Manual ...
Page 69: ...Vertex Edge System 68 User Manual ...
Page 91: ...Vertex Edge System 90 User Manual ...
Page 96: ...Vertex Edge System 95 User Manual ...
Page 98: ...1 Main power ON OFF switch 2 Analyzer power switches Vertex Edge System 97 User Manual ...
Page 165: ... September 2021 1998 2002_EN_RevA1_VertexEdge_UserManual Vertex Edge System 164 User Manual ...