2.
Locate the Ethernet cable on the rear panel and disconnect it.
3.
Open the side of the Analyzer by unscrewing the 4 thumbscrews at the top of the left side.
4.
Approximately 20 seconds after the Ethernet Cable has been disconnected, the Analyzer
will recognize that it has lost communications with the DAq and activate the “Manual
Override” button. The LED’s on the front of the Analyzer will flash to show a Maintenance
Fault per the LED status flash pattern.
To reset faults and alarms press and hold button for 1-3 seconds.
To Exit Monitor and Open Gate press and hold button for 10 seconds and above
To restart monitoring mode, press and hold button for 4 to 9 seconds.
5.
Close and reattach the Analyzer cover, reconnect the Ethernet Communications Cable
and slide the Analyzer into the cabinet.
Fix an Unresponsive Vertex Edge Touch Screen
Completely resetting the computer can resolve many issues that cause a frozen/unresponsive
screen. Follow these steps to perform a hard reset:
1.
Remove any USB devices from the USB ports of the HMI PC if non-default USB devices
were inserted. The USB connection from touch screen should remain.
2.
Turn off the computer by pressing and holding the power button for a few seconds.
Vertex Edge System
142
User Manual
Summary of Contents for Vertex Edge
Page 1: ...Vertex Edge 72 POINT CONTINUOUS MONITOR ...
Page 2: ......
Page 6: ......
Page 19: ...Chemcassette 1 Chemcassette directional flow Vertex Edge System 18 User Manual ...
Page 29: ...Installation Drawings Vertex Edge System 28 User Manual ...
Page 46: ......
Page 59: ...Vertex Edge System 58 User Manual ...
Page 63: ...Vertex Edge System 62 User Manual ...
Page 69: ...Vertex Edge System 68 User Manual ...
Page 91: ...Vertex Edge System 90 User Manual ...
Page 96: ...Vertex Edge System 95 User Manual ...
Page 98: ...1 Main power ON OFF switch 2 Analyzer power switches Vertex Edge System 97 User Manual ...
Page 165: ... September 2021 1998 2002_EN_RevA1_VertexEdge_UserManual Vertex Edge System 164 User Manual ...