Instrument Faults
An instrument fault indicates a loss of monitoring on one or more points.
Event
ID
Description
Possible Cause
Resolution
201
Chemcassette tape decolored
Improper Chemcassette storage
Confirm Chemcassette storage meeting temperature meets or exceeds the
Chemcassette storage requirements
Replace Chemcassette
Tape is too old
Confirm Expiration date will reach the tape length time line
Replace Chemcassette
202
End Of Chemcassette
End of Chemcassette
Replace Chemcassette
Chemcassette broken
Rethread Chemcassette
Faulty tape encoder
Service analyzer
Gate opening insufficient
Service analyzer
203
Failed writing hardware configuration
Non-volatile memory failure in analyzer CPU
Retry and reboot the Analyzer.
Service or replace analyzer.
204
Failed reading hardware configuration
Non-volatile memory failure in analyzer CPU
Retry and reboot the Analyzer.
Service or replace analyzer.
205
Gate motor times out
Gate position sensor not activated before timeout
Check motor operation using Maintenance, Analyzer Operations, Open Gate, or
Close Gate
Motor does not operate
Check motor connections Analyzer PCB in analyzer
Bad sensor or cable
Check sensor connection to Analyzer PCB.
Contact Honeywell service.
206
Gate motor driving failure
Gate position sensor not activated before timeout
Check motor operation using Maintenance, Analyzer Operations, Open Gate, or
Close Gate
Motor does not operate
Check motor connections to Analyzer PCB in analyzer
Bad sensor or cable
Check sensor connection to Analyzer PCB. Contact Honeywell service
207
Optics signals are noisy
Poor grounding
Replace Analyzer, Contact Honeywell service
Optics block cover loose
Retighten or reinstall as required
208
Optics counts very low <200
Optics LED not properly calibrated
Perform Replace CC or Adjust Optics operation to recalibrate
Cable disconnected
Check cable between Optic Block and Analyzer PCB.
Contact Honeywell service.
Optics board defective
Replace the optics block. Contact Honeywell service
Analyzer CPU defective
Replace Analyzer, Contact Honeywell service
209
Gas table file is bad or missing
No configuration loaded
Reinstall Configuration Profile
Vertex Edge System
133
User Manual
Summary of Contents for Vertex Edge
Page 1: ...Vertex Edge 72 POINT CONTINUOUS MONITOR ...
Page 2: ......
Page 6: ......
Page 19: ...Chemcassette 1 Chemcassette directional flow Vertex Edge System 18 User Manual ...
Page 29: ...Installation Drawings Vertex Edge System 28 User Manual ...
Page 46: ......
Page 59: ...Vertex Edge System 58 User Manual ...
Page 63: ...Vertex Edge System 62 User Manual ...
Page 69: ...Vertex Edge System 68 User Manual ...
Page 91: ...Vertex Edge System 90 User Manual ...
Page 96: ...Vertex Edge System 95 User Manual ...
Page 98: ...1 Main power ON OFF switch 2 Analyzer power switches Vertex Edge System 97 User Manual ...
Page 165: ... September 2021 1998 2002_EN_RevA1_VertexEdge_UserManual Vertex Edge System 164 User Manual ...