available
…
…
40008
Pt 8 Status
Unsigned
Integer
1
40009
Az Status
Unsigned
Integer
1
Analyzer overall status l Bit 0: AZ enable status (0: disabled, 1: enabled) l Bit 1: monitoring status (0: idle, 1: in monitoring mode) l Bit 2: Alarm 1 (0: normal, 1: in
gas alarm) l Bit 3: Alarm 2 (0: normal, 1: in gas alarm) l Bit 4: latched alarm 1 (0: normal, 1: in gas alarm) l Bit 5: latched alarm 2 (0: normal, 1: in gas alarm) l Bit 6:
MFault (0: normal, 1: in Mfault) l Bit 7: IFault (0: normal, 1: in Ifault) l Bit 8: Sim Alarm 1 (0: normal, 1: in gas alarm) l Bit 9: Sim Alarm 2 (0: normal, 1: in gas alarm)
l Bit 10: Sim IFault (0: normal, 1: in Ifault) l Bit 11: Sim MFault (0: normal, 1: in Mfault) l l Bit 13: Reserved l Bit 14: Reserved l Bit 15: Reserved
40010
Az Flt
Heartbeat
unsigned
integer
1
0: No fault 1: Maintenance Fault 2: Instrument Fault 3: Both faults present A heartbeat counter is provided with some of the fieldbuses in word 9. This
increments every second to confirm that communications is working
40011
Pt 1 GasConc
Floating
point
2
…
…
…
…
40025
Pt 8 GasConc
Floating
point
2
40027
Pt 1 NormalizedConc
Unsigned
Integer
1
Not used for now
…
…
…
…
40034
Pt 8 NormalizedConc
Unsigned
Integer
1
40035
Pt 1 Flow
Unsigned
Integer
1
cc/min
…
…
40042
Pt 8 Flow
Unsigned
Integer
1
40043
Az Optics1 Status
Unsigned
integer
1
Optic Status 5th, 6th byte in optic communication [Low byte] 0x00 No faults 0x01 Watchdog reset 0x02 SW Error (over-range, pointer, buffer overflow, switch-
case) 0x04 ADC conversion overflow 0x08 ADC timing overflow 0x10 Post-processing overrun [High byte] 0x00 No faults 0x01 NVM CRC mismatch (at power
up and at each access) 0x02 Program CRC mismatch (at power up) 0x04 SPI protocol error 0x08 SPI Tx/Rx overrun 0x10 SPI ISR timeout 0x20 SPI APP timeout
40044
Az Optics2 Status
Unsigned
integer
1
40045
Az Pump1
Status
Az Pump2
Status
1
0: Unknown, 1: Bad, 2: good, 3: Service required
40046
Az active fault code
unsigned
integer
1
if no fault, then fault code should be 0x0 If there are active multiple instrument faults, the lowest instrument fault will be returned If there are both instrument
and maintenance faults, the lowest instrument fault will be returned
40047
Az CCDays
signed
integer
1
40048
Az FilterLife
signed
integer
1
40049
Az PumpMaintDays
signed
1
Vertex Edge System
156
User Manual
Summary of Contents for Vertex Edge
Page 1: ...Vertex Edge 72 POINT CONTINUOUS MONITOR ...
Page 2: ......
Page 6: ......
Page 19: ...Chemcassette 1 Chemcassette directional flow Vertex Edge System 18 User Manual ...
Page 29: ...Installation Drawings Vertex Edge System 28 User Manual ...
Page 46: ......
Page 59: ...Vertex Edge System 58 User Manual ...
Page 63: ...Vertex Edge System 62 User Manual ...
Page 69: ...Vertex Edge System 68 User Manual ...
Page 91: ...Vertex Edge System 90 User Manual ...
Page 96: ...Vertex Edge System 95 User Manual ...
Page 98: ...1 Main power ON OFF switch 2 Analyzer power switches Vertex Edge System 97 User Manual ...
Page 165: ... September 2021 1998 2002_EN_RevA1_VertexEdge_UserManual Vertex Edge System 164 User Manual ...