High Voltage Probe 

HV-P60A 

 
 
 
 
 
 
 
 
 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Instruction Manual

 

 

Summary of Contents for HV-P30A

Page 1: ...High Voltage Probe HV P60A Instruction Manual ...

Page 2: ...Ⓒ2015 IWATSU TEST INSTRUMENTS CORPORATION All rights reserved ...

Page 3: ... user or damage to property read and carefully observe warnings and cautions in the following sections Definition of warnings and cautions used in this manual Warnings Incorrect operation or malfunction to observe the warning may result in death or serious injury Cautions Incorrect operation or malfunction to observe the caution may result in injury or damage to the instrument Notes Parts of the c...

Page 4: ...he grouding lead is brought into contact with the insulator or nose of the probe it may result in an electric shock fire or malfunction Coating of the grouding lead is voltage resistant approx 10 kV but a clip is not and should be used carefully Do not apply a voltage higher than the specified one If a voltage higher than the specified is applied it may result in an electric shock fire or malfunct...

Page 5: ... or malfunction If water is allowed in it immediately turn off the power switch of the object to be measured and remove this probe from the object to be measured after confirming the thing that a high voltage has not gone out and contact Iwatsu office or our sales distributors for repair Do not use this instrument when the drop of water adheres to it or the hand is wet Otherwise electric shock fir...

Page 6: ... to it A strong shock may damage the instrument and result in malfunction or an electronic shock When this instrument is dropped or damaged contact Iwatsu office or our sales distributors for repair Do not modify or disassemble this instrument The repair and remodeling by the customer cause an electric shock fire or malfunction The user cannot repair this instrument Contact Iwatsu office or our sa...

Page 7: ...er begins pull out the power plug of the target or the oscilloscope to which this instrument is connected from the outlet Then detach this instrument from the oscilloscope or the object to be measured and do not use it If not lightning may cause an electric shock fire or malfunction When the insulator is contaminated wipe off contamination prior to using it If the insulator is contaminated a high ...

Page 8: ...ed from this instrument before maintenance In addition wipe off with a dry cloth when water droplets are attached If not an electric shock or malfunction may occur Before transporting this instrument pack the housed instrument in the packing material used at the time of purchase or the equivalent or better This instrument may fail if a large vibration or impact is applied during transportation pos...

Page 9: ...ls while referring to the following Components If there is a missing item or an item damaged during transportation immediately contact Iwatsu office or our sales distributors Components Probe body 1 Accessories Output cable 1 Clamp 1 Tip 1 Ground lead 1 Adjusting screwdriver 1 Carrying bag 1 Instruction Manual 1 ...

Page 10: ...s instrument description of the malfunction and the name division and telephone number of the responsible person Use the box with a packing box used when goods is delivered or an equal cushioning material to avoid the accident in transit when you send it If you do not have a proper packing or shock absorbing material for transportation contact Iwatsu office or our sales distributors Cleaning of th...

Page 11: ...aired viii Cleaning of this instrument viii Storage viii 1 General Description 2 2 Name of each part and composition 2 2 1 Probe body and Accessories 2 2 2 Carrying bag accessory 3 3 Adjustment of Probe 4 3 1 Preparations 4 3 2 Adjustment 5 4 Measuring method 7 4 1 Installation and connection 7 4 2 Measuring Procedure 12 5 Specifications 13 5 1 Instrument specifications 13 5 2 Appearance 15 Contac...

Page 12: ...y bandwidth is DC to 50 MHz maximum input voltage is 60 Kv and attenation ratio is 2000 1 2 Name of each part and composition 2 1 Probe body and Accessories This instrument is composed of a probe body of Figure 1 The output cable is contained and various accessories Figure 1 Probe body and Accessories 付属品 Screwdriver Probe body Tip Clamp Ground lead Matching box part ...

Page 13: ...essory A probe body and various accessories are bundled to Carrying bag like Figure 2 and delivered Moreover the probe body and the accessory can be carried by putting it in Carrying bag after packing things are opened Figure 2 Carrying bag accessory ...

Page 14: ...e The waveform of this probe is adjusted with the following preparations according to the procedure 3 1 Preparations Prepare the signal generator that can generate the following signal specifications and the measuring instrument oscilloscope used for the measurement thereafter As for the example of the used equipment it is not the following Signal generator Waveform Square wave Frequency 1 kHz or ...

Page 15: ... INPUT of the oscilloscope is connected with the connector of the probe body with Output cable 3 The output of the signal generator is connected with the tip of the probe body Signal Generator Figure 3 a Connection when probe is adjusted Oscilloscope Probe body Ground terminal Matching Box Connector Tip of Probe Figure 3 b Waveform adjustment of probe in matching box closeup Screwdriver Hole for w...

Page 16: ... the adjustment is inserted in the adjustment hole of the Matching box Refer to Figure 3 b The waveform of the probe is adjusted to become the ideal waveform like Figure 3 e Adequated while seeing the waveform following Figure 3 c and Figure 3 d displayed in the oscilloscope Figure 3 d Undercompensated Figure 3 c Overcompensated Figure 3 e Adequated ...

Page 17: ... the hole plug on the stand side and it is drawn out through the hole in the probe receiving part 3 The output cable is connected with the bottom of Probe body 4 Probe body is inserted in the probe receiving part and is tightened with two fixed screws to be fixed 5 Please connect the terminal GND of the probe stand with earth GND for safety Figure 4 a Probe body HV P60A and Probe stand SK 301 Prob...

Page 18: ...nd Figure 4 a on the stand side and it is drawn out from the hole in the probe receiving part 4 The output cable is connected with the bottom of Probe body 5 The probe receiving part is fixed to the stand with four make up screws Refer to Figure 4 b 6 The position in which the stand is installed is chosen and the stand installation hole is fixed with the screw etc Refer to four places and Figures ...

Page 19: ... High Voltage Probe HV P60A Figure 4 c Screw hole position for installation of Probe body on Probe receiving part Probe receiving part back Screw hole position for installation of Probe body four places ...

Page 20: ...forehand 3 The connector of Probe body and INPUT of the oscilloscope is connected with the output cable Connect the output cable with INPUT of the oscilloscope when the connector of Probe body is connected with the output cable beforehand by 1 on the page beforehand 4 The earth lead of the attachment is touched to the measurement part beforehand and it is confirmed the voltage is not generated Cau...

Page 21: ... measured and connected with Tip of Probe Refer to Figure 6 When connecting the lead wire it is tightened firmly with Clamp of the attachment Figure 6 Connection of lead wire for measurement Probe body Connector Output cable Oscilloscope Matching box Ground lead Ground terminal Object to be measured Clamp Tip Measurement part Lead wire for measurement Earth lead of the attachment ...

Page 22: ... in addition to the earth lead of the attachment to decrease ringing and the overshoot Prepare the metal with high electric conductivity treatment device tie to the earth lead joint and strengthen GND In that case do not touch the probe point with GND As a concrete example there are the following methods of etc 1 The stick made of copper about φ5 mm is used 2 The thin sheet copper is folded and us...

Page 23: ...om the next expression Tr ns 350 Band width MHz Attenuation ratio 2000 1 5 Waveform distortion 5 10 to 35 Note Ringing and Overshoot are excluded Input RC R 1000 MΩ 5 C 5 pF 2 pF Output cable Approx 4 m Connector type BNC type Input RC of oscilloscope that can be connected R 1MΩ C 10 pF to 40 pF Dimensions Refer to 5 2 Appearance Weight Probe body Approx 2 0 kg Note Output cable and Matching box a...

Page 24: ...14 High Voltage Probe HV P60A Maximum input voltage DC Peak AC ...

Page 25: ...15 High Voltage Probe HV P60A 5 2 Appearance Unit mm ...

Page 26: ...Address 7 41 Kugayama 1 chome Suginami ku Tokyo 168 8511 Japan Phone 81 3 5370 5483 Facsimile 81 3 5370 5492 Homepage http www iti iwatsu co jp ...

Page 27: ...HV P60A ...