v
High Voltage Probe HV-P60A
Read the following safety information
Warnings
●
Connect the playground of the probe body with the grounding
potential (playground) of the object to be measured.
When the playground of the probe body is connected with potential other
than the playground of the object to be measured, it might cause an
electric shock or malfunction (for the object to be measured, this
instrument, and connected oscilloscope).
●
When thunder begins, pull out the power plug of the target or
the oscilloscope to which this instrument is connected from
the outlet. Then, detach this instrument from the oscilloscope
or the object to be measured, and do not use it.
If not, lightning may cause an electric shock, fire, or malfunction.
●
When the insulator is contaminated, wipe off contamination
prior to using it.
If the insulator is contaminated, a high voltage may leak and cause an
electric shock, fire or malfunction. Apply a little amount of neutral detergent
to soft cloth and wipe the insulator gently with it. Do not use highly volatile
liquid such as benzene or thinner.