High Voltage Probe HV-P60A
(2) Connection of oscilloscope and object to be measured
(Refer to Figure 5.)
1) The power switch of the object to be measured is turned off.
2) The ground terminal of Probe body is connected with the earth
GND of the object to be measured and the terminal GND of the
oscilloscope.
Connect the terminal GND of the stand with the earth of the
object to be measured and the terminal GND of the oscilloscope
when connecting the ground terminal of Probe body with the
terminal GND of the stand by (1) on the page beforehand.
3) The connector of Probe body and INPUT of the oscilloscope is
connected with the output cable.
Connect the output cable with INPUT of the oscilloscope when
the connector of Probe body is connected with the output cable
beforehand by (1) on the page beforehand.
4) The earth lead of the attachment is touched to the measurement
part beforehand, and it is confirmed the voltage is not generated.
Caution
!
Connect following 2), 3) and 5) with each terminal firmly so that
there is no shake.
Probe body
Figure 5
Example of connecting measurement
(object to be measured and oscilloscope)
Ground terminal
Object to be
measured
Oscilloscope
Matching box
Earth lead of the attachment
Ou
tp
u
t
ca
b
le
Ground lead
Connector
Clamp
Tip
Measurement part
Summary of Contents for HV-P30A
Page 1: ...High Voltage Probe HV P60A Instruction Manual ...
Page 2: ...Ⓒ2015 IWATSU TEST INSTRUMENTS CORPORATION All rights reserved ...
Page 24: ...14 High Voltage Probe HV P60A Maximum input voltage DC Peak AC ...
Page 25: ...15 High Voltage Probe HV P60A 5 2 Appearance Unit mm ...
Page 27: ...HV P60A ...