Orifice Filter Inspection, Cleaning & Replacement
l
The orifice is protected by a mesh filter.
In the event that a particle or foreign object makes it past the end of sample line filter this
mesh filter will protect the orifice from clogging and slowing down sample transport
times.
l
This should be inspected every 6 months for debris or contamination and can be cleaned
with compressed air.
l
In the event that the filter can not be cleaned it should be replaced.
l
Replacement Part: 1502A0155
NOTE
This inspection should only take place on analyzers that are disconnected from the rack, or
if the ENTIRE rack is out of monitor mode.
1.
Release the Analyzer from the rack (refer to See "Release the Analyzer Lock" on page
2.
Remove the orifice Panel on the side of the Analyzer by unscrewing the thumbscrew
3.
Thread in the orifice removal tool 1502-0166 or a 6-32 screw into the orifice cartridge
and pull back to remove it from the manifold
Vertex Edge System
118
User Manual
Summary of Contents for Vertex Edge
Page 1: ...Vertex Edge 72 POINT CONTINUOUS MONITOR ...
Page 2: ......
Page 6: ......
Page 19: ...Chemcassette 1 Chemcassette directional flow Vertex Edge System 18 User Manual ...
Page 29: ...Installation Drawings Vertex Edge System 28 User Manual ...
Page 46: ......
Page 59: ...Vertex Edge System 58 User Manual ...
Page 63: ...Vertex Edge System 62 User Manual ...
Page 69: ...Vertex Edge System 68 User Manual ...
Page 91: ...Vertex Edge System 90 User Manual ...
Page 96: ...Vertex Edge System 95 User Manual ...
Page 98: ...1 Main power ON OFF switch 2 Analyzer power switches Vertex Edge System 97 User Manual ...
Page 165: ... September 2021 1998 2002_EN_RevA1_VertexEdge_UserManual Vertex Edge System 164 User Manual ...