Family
CC Name (P/N)
Table
Gas Name
TLV
LAL
LDL
Default
Alarm
Level 1
Default
Alarm
Level 2
Range
Alarm
Setting
Initial
Analysis
Period
(second)
Time to 1TLV alarm @
2TLV concentration, 10ft
sample line
(Low Level)
ppm
0.2-1.0 ppm
60
with 0.2ppmF2 gas)
3
Chlorine XP-Cl2-II (Cl2)
0.1 ppm
0.06 ppm
0.05
ppm
0.5 ppm
1 ppm
0-5 ppm
0.06-0.24 ppm
0.25-5.0 ppm
45
30
<40sec (Alarm @ 0.1ppm
with 0.2ppmCl2 gas).
<20sec (Alarm @ 0.5ppm
with 1.0ppmCl2 gas)
4
Chlorine XP-Cl2-II (Cl2)
(Low Level)
0.1 ppm
0.03 ppm
0.007
ppm
0.1 ppm
0.2 ppm
0-1.0 ppm
0.03 - 0.099 ppm
0.1 - 0.199 ppm
0.2 -1.0 ppm
120
90
60
<110 sec (Alarm @ 0.1ppm
with 0.2ppmCl2 gas)
12
Fluorine/Oxidizer
(1295-0220)
1
Chlorine (Cl2)
0.1 ppm
0.05 ppm
0.04
ppm
0.5 ppm
1 ppm
0-5 ppm
0.05-0.24 ppm
0.25-0.49 ppm
0.5-5 ppm
45
30
15
<40 sec (Alarm @ 0.5ppm
with 1ppm Cl2 gas)
2
Fluorine (F2)
0.1 ppm
0.1 ppm
0.06
ppm
1 ppm
2 ppm
0-10 ppm
0.1-0.9 ppm
1.0-10 ppm
60
30
<80 sec (Alarm @ 0.1ppm
with 0.2 ppmF2 gas)
3
Nitrogen Dioxide (NO2)
0.2 ppm
0.1 ppm
0.1 ppm 0.2 ppm
0.4 ppm
0-30 ppm
0.1-8.9 ppm
9-30 ppm
240
120
<350 sec
13
Sulfur Dioxide (1295-
0552)
1
Sulfur Dioxide (SO2)
250 ppb
30 ppb
25 ppb
250 ppb
500 ppb
0-2500 ppb
30-249 ppb
250-2500 ppb
60
30
<60 sec
14
Hydrogen Cyanide
(1295-0222)
1
Hydrogen Cyanide
(HCN)
4.7 ppm
1 ppm
0.5 ppm 4.7 ppm
9.4 ppm
0-30 ppm
1-9.9 ppm
10-19.9 ppm
20-30 ppm
30
20
15
<15 sec
15
XP4 Chlorine (1257-
9308)
1
Chlorine (Cl2)
0.1ppm
0.05 ppm
0.05
ppm
0.5 ppm
1.0 ppm
0-5 ppm
0.05-0.24 ppm
0.25-5.0 ppm
45
30
<53 sec
2
Chlorine (Cl2) (Low
Level)
100 ppb
30 ppb
7 ppb
100 ppb
200 ppb
0 - 2000
ppb
30 - 199 ppb
200 - 499 ppb
500 -2000 ppb
120
90
60
<65 sec
TLV - Threshold Limit Value — LAL - Lowest Alarm Level — LDL - Low Detection Limit
The response time (T50) generally indicates the amounts of time required to display 50% of the accurate reading of the actual gas concentration, determined under standard Lab condition (20-25 C, 42-46%RH) at
the standard analysis point flow rate 200cc/min. The performance efficiency including response accuracy and response time can deviate depending on sample gas conditions. (ex. The response time will be extended
at higher RH condition due to absorption onto the sample line.)
Vertex Edge System
129
User Manual
Summary of Contents for Vertex Edge
Page 1: ...Vertex Edge 72 POINT CONTINUOUS MONITOR ...
Page 2: ......
Page 6: ......
Page 19: ...Chemcassette 1 Chemcassette directional flow Vertex Edge System 18 User Manual ...
Page 29: ...Installation Drawings Vertex Edge System 28 User Manual ...
Page 46: ......
Page 59: ...Vertex Edge System 58 User Manual ...
Page 63: ...Vertex Edge System 62 User Manual ...
Page 69: ...Vertex Edge System 68 User Manual ...
Page 91: ...Vertex Edge System 90 User Manual ...
Page 96: ...Vertex Edge System 95 User Manual ...
Page 98: ...1 Main power ON OFF switch 2 Analyzer power switches Vertex Edge System 97 User Manual ...
Page 165: ... September 2021 1998 2002_EN_RevA1_VertexEdge_UserManual Vertex Edge System 164 User Manual ...