Replace the Chemcassette
Change the Vertex Edge Chemcassette tape for any of the following reasons:
l
Scheduled end-of-tape service
l
Low Chemcassette warning
l
Chemcassette has expired
l
End of Chemcassette
1. In the left navigation panel, tap OVERVIEW
2. Tap on the selected Analyzer.
3. In the Components status bar,
tap on the Chemcassette button
.
4. In the Chemcassette window, tap REPLACE CC.
5. “Stopping monitoring mode” confirmation screen pops up. Tap CONFIRM.
6. The tape replacement procedure gets started. The Analyzer is released, and the Optics gate
opens.
7. Pull out the Analyzer and remove old Chemcassette tape.
Vertex Edge System
99
User Manual