Symbol
Gas Name
Filter Type
A
Filter Type
B
Filter Type
C
HCN
Hydrogen Cyanide
X
HF
Hydrogen Fluoride
X
X
HF-LL
Hydrogen Fluoride - Low Level
X
X
H
2
Se
Hydrogen Selenide
X
NO
2
Nitrogen Dioxide
X
COCI
2
Phosgene
X
COCI
2
/-
HL
Phosgene - High Range
X
PH
3
Phosphine
X
SiH
4
Silane
X
SO
2
Sulful Dioxide
X
X
TDMAT
Tetrakis Dimethylamino
Titanium
X
X
Nominal Transport Times
The following table shows the time required for samples to move from the sampling point to the
Vertex Edge System for various lengths of sample lines.
OD
0.25" Outside Diameter Tubing
ID
0.190"(Thin Wall) ID
Length in feet
100
200
300
327
400
Length in meters
30
61
91
100
122
Sample flow
1.6
20.5
40.9
61.4
66.9
81.9
2.1
15.6
31.2
46.8
51.0
62.4
2.2
14.9
29.8
44.7
48.7
59.5
2.4
13.6
27.3
40.9
44.6
54.6
Vertex Edge System
145
User Manual
Summary of Contents for Vertex Edge
Page 1: ...Vertex Edge 72 POINT CONTINUOUS MONITOR ...
Page 2: ......
Page 6: ......
Page 19: ...Chemcassette 1 Chemcassette directional flow Vertex Edge System 18 User Manual ...
Page 29: ...Installation Drawings Vertex Edge System 28 User Manual ...
Page 46: ......
Page 59: ...Vertex Edge System 58 User Manual ...
Page 63: ...Vertex Edge System 62 User Manual ...
Page 69: ...Vertex Edge System 68 User Manual ...
Page 91: ...Vertex Edge System 90 User Manual ...
Page 96: ...Vertex Edge System 95 User Manual ...
Page 98: ...1 Main power ON OFF switch 2 Analyzer power switches Vertex Edge System 97 User Manual ...
Page 165: ... September 2021 1998 2002_EN_RevA1_VertexEdge_UserManual Vertex Edge System 164 User Manual ...