Event
ID
Description
Possible Cause
Resolution
233
Tape advance failure
End of Chemcassette
Replace Chemcassette
Encode roller not rotating due to loose tape
Reload the Chemcassette tape
Encoder failure
Replace Encoder
Cable issue
Check Encoder cabling
Contact Honeywell Analytics Service
Stepper Motor Failure
Replace Stepper Motor
Stepper Motor Cable issue
Check Stepper Motor cabling
Contact Honeywell Analytics Service
234
Internal Comm failure
Hardware failure
Contact Honeywell Analytics Service
235
Abnormal Az current consumption is
detected
Hardware failure
Contact Honeywell Analytics Service
236
Internal fault at control module
Hardware failure
Contact Honeywell Analytics Service
237
Internal fault at PDU module
Hardware failure
Contact Honeywell Analytics Service
238
Analyzer SW corrupted
Software installation failure
Re-install analyzer software
Replace Analyzer, Contact Honeywell service
239
Internal SW fault at analyzer
Internal SW operation failed
Reboot the Analyzer. Contact Honeywell service.
Vertex Edge System
136
User Manual
Summary of Contents for Vertex Edge
Page 1: ...Vertex Edge 72 POINT CONTINUOUS MONITOR ...
Page 2: ......
Page 6: ......
Page 19: ...Chemcassette 1 Chemcassette directional flow Vertex Edge System 18 User Manual ...
Page 29: ...Installation Drawings Vertex Edge System 28 User Manual ...
Page 46: ......
Page 59: ...Vertex Edge System 58 User Manual ...
Page 63: ...Vertex Edge System 62 User Manual ...
Page 69: ...Vertex Edge System 68 User Manual ...
Page 91: ...Vertex Edge System 90 User Manual ...
Page 96: ...Vertex Edge System 95 User Manual ...
Page 98: ...1 Main power ON OFF switch 2 Analyzer power switches Vertex Edge System 97 User Manual ...
Page 165: ... September 2021 1998 2002_EN_RevA1_VertexEdge_UserManual Vertex Edge System 164 User Manual ...