3.
If replacing Pump 1, disconnect the power connector before opening drawer. (Pump 2
power connector can be removed with the drawer opened). Power connector removal are
in steps 4-5.
4.
Uninstall the electrical connector on the side of the pump
a.
Pull the white tab out
b.
Push down on the black tab and pull back on the connector
Vertex Edge System
110
User Manual