Replacing the Sensor (PC11 Sensor), and (PC20 Sensor)
Tools Needed: Magnetic Philips Screwdriver (No.2)
①
Remove the Main Cover, the Side Cover, and the Back Cover. For instructions, please refer to “8.1
Preparation”
②
Remove the Guide (Lower Mirror), the Cover Sheet. For instructions, please refer to “8.3.13 Lower Mirror
ASSY”
③
Unscrew 6 screws on the Upper Guide ASSY, then remove the Upper Guide ASSY
H8-37
「■
Replacing the Prism (PC21 Sensor), the PC 22 ASSY (PC22 Sensor), the Prism (PC31 Sensor)
」
④
Remove the Pick Unit ASSY
⑤
Remove the Pick Cover (U)
⑥
Loosen the screws on the Ultra Sonic Sensor
⑦
Unscrew 4 bind-head screws on the M2-ASSY Cover, then remove the M2-ASSY Cover
M2-ASSY Cover
Bind-head screw
Bind-head screw
Bind-head screw
Bind-head screw
⑧
Unscrew 2 screws on the D1 Pinch Roller, and remove the Pinch Roller
⑨
Remove the Sensor
- Flip the M2-ASSY Cover and remove the clips on the Sensor (PC11 Sensor) and remove the Sensor
- Flip the M2-ASSY Cover and remove the clips on the Sensor (PC20 Sensor) and remove the Sensor
⑩
Attach the new Sensor
⑪
Take the reverse steps to attach the Sensor
⑫
After replacing the part, run Auto Calibration Test
#55 Sensor
(PC11 emission)
#55 Sensor (PC20 emission)
M2-ASSY Cover
Caution
:
:
:
:
When removing the Upper Guide ASSY, disconnect the PC22 cable connector, and
unlock the locking wire saddle.
Summary of Contents for HT-4139-28
Page 1: ...HT 4139 28 48 Scanner Maintenance Manual ...
Page 2: ......
Page 11: ... H2 5 2 5 Block Diagram ...
Page 22: ...REV0 10 02 15 H3 10 Lens Mirror Motor Mirror No 1 Mirror No 3 Mirror No 2 ...
Page 35: ...REV0 10 02 15 H7 2 ...
Page 63: ...REV0 2010 02 15 H7 30 ...
Page 89: ...REV0 10 02 15 H8 26 ...