47
Connecting Measurement Leads and Sensors to Lines to Be Measured
2.8 Connecting Measurement Leads and Sensors to
Lines to Be Measured
Perform zero adjustment, and then connect voltage cords and current sensors to lines to be
measured as indicated in the wiring diagram shown on
[INPUT]
> the
[WIRING]
screen. To ensure
accurate measurement, connect the instrument exactly as shown on
[INPUT]
> the
[WIRING]
screen.
The wiring diagram will be displayed when you select a wiring mode on
[INPUT]
> the
[WIRING]
screen.
See “2.5 Setting Wiring Mode and Configuring Current Sensor Settings” (p. 41).
Voltage Cord
Firmly clip the voltage cords to the metallic parts on the power supply
side, such as screws and bus bars
Current sensor
Source
Load
Clamp the current sensor so that its current direction mark points at the
load side.
OK
NO
NO
NO
Do not clamp the sensor
around two or more conductors.
Do not pinch the conductor. Do not clamp the
sensor to a shielded
wire.
IMPORTANT
•
The phases are labeled as
A
,
B
,
C
on the wiring diagram screen. Connect the instrument based
on whatever names you are using, such as
R/S/T
and
U/V/W
, as appropriate.
•
Clamp the sensor around only one conductor. Clamping the sensor around two or more of
conductors in a bundle prevents the instrument from measuring any current regardless of
whether the measurement target is a single-phase or three-phase circuit.
Preparing for Measurement
HIOKI PW8001A961-00
Summary of Contents for PW8001-01
Page 2: ...HIOKI PW8001A961 00...
Page 6: ...Contents iv HIOKI PW8001A961 00...
Page 56: ...50 Checking Connections HIOKI PW8001A961 00...
Page 120: ...114 Power Calculation Method HIOKI PW8001A961 00...
Page 162: ...156 Controlling Integration with External Signals HIOKI PW8001A961 00...
Page 236: ...230 U7005 15MS s Input Unit HIOKI PW8001A961 00...
Page 258: ...252 Index HIOKI PW8001A961 00...
Page 259: ...HIOKI PW8001A961 00...
Page 260: ...HIOKI PW8001A961 00...
Page 261: ...HIOKI PW8001A961 00...
Page 262: ...www hioki com HIOKI PW8001A961 00...