84
Motor Measurement (Motor Analysis-Equipped Model)
Configuring the motor input settings
Connect torque sensors and tachometers by referring to “Motor measurement wiring” (p. 78).
Configure the motor analysis settings based on those connections.
Display screen
[INPUT]
>
[MOTOR]
1
1
3
3
4
4
5
5
6
6
2
2
1
Tap
[Motor analysis option wiring]
to
select.
2
Tap
[CH A-D]
or
[CH E-H]
to display
channels you wish to change the
settings.
3
Tap the
[Upper
f
lim.]
and
[Lower
f
lim.]
box, and select the frequency from the
list.
Set this when pulses are to be input for the
motor input.
Upper
f
lim.
100 Hz, 500 Hz, 1 kHz, 5 kHz,
10 kHz, 50 kHz, 100 kHz,
500 kHz, 1 MHz, 2 MHz
Lower
f
lim.
0.1 Hz, 1 Hz, 10 Hz, 100 Hz
Upper frequency limit
This setting specifies the lowest frequency that exceeds the maximum frequency of the input pulse signal.
When
[Motor analysis option wiring]
is set to
[Individual Input]
, the setting is used as the upper limit for
D/A output.
When using motor analysis mode, this setting is used as the pulse frequency that is used to display RPMs and
motor powers and as a pulse frequency to calculate the upper limit value for the D/A output.
(RPM upper limit value) =
60
×
(Set upper frequency limit)
(Pulse count setting)
(Motor power upper limit value) = (Maximum torque value)
×
2
×
π
×
(RPM upper limit value)
60
If the input RMS signal setting is
[Analog]
, the RMS upper limit is calculated by multiplying the scaled RMS
value by the voltage range value.
Lower frequency limit
This setting specifies the lower frequency limit for measurement of the input pulse signal.
When the following synchronization sources are selected, the lower limit frequency is also used as the lower
frequency limit for measurement.
Ext1, Ext2, Ext3, Ext4
Zph1, Zph3
Ch. B, Ch. D, Ch. F, Ch. H
HIOKI PW8001A961-00
Summary of Contents for PW8001-01
Page 2: ...HIOKI PW8001A961 00...
Page 6: ...Contents iv HIOKI PW8001A961 00...
Page 56: ...50 Checking Connections HIOKI PW8001A961 00...
Page 120: ...114 Power Calculation Method HIOKI PW8001A961 00...
Page 162: ...156 Controlling Integration with External Signals HIOKI PW8001A961 00...
Page 236: ...230 U7005 15MS s Input Unit HIOKI PW8001A961 00...
Page 258: ...252 Index HIOKI PW8001A961 00...
Page 259: ...HIOKI PW8001A961 00...
Page 260: ...HIOKI PW8001A961 00...
Page 261: ...HIOKI PW8001A961 00...
Page 262: ...www hioki com HIOKI PW8001A961 00...