110
Peak Hold Function
Operation in the peak hold state
• Peak hold operation also applies to the following measured values:
(1) Measured values to be stored onto the USB flash drive
(2) Measured values to be acquired using communications
(3) Measured values to be outputted as an analog signal
• Waveforms, the clock, and the peak-over display are refreshed.
• When the display is overloaded, it will show
[------]
. In this case, cancel the peak hold operation,
then switch the present range over to a range in which the overload condition will not occur.
• The maximum value is determined using the absolute value of measured values. (However, this
method does not affect voltage or current peak values.)
•
For example, if a power of −60 W is inputted after that of 50 W was inputted, the display will
indicate
[−60 W]
because the absolute value of
−60 W
is greater than that of
50 W
.
• When the
HOLD
key is pressed, the peak hold value is reset, and another peak hold operation
starts from that point.
• The instrument retains the peak-hold value without updating it even if the interval time set in the
time control function has elapsed.
• During average calculation, peak hold operation applies to measured values after averaging.
•
Settings that affect measured values such as range and the LPF settings cannot be changed.
• When the range setting is set to
[AUTO]
, the range is fixed to that at the time when the
PEAK HOLD
key was pressed.
• The hold function cannot be used together with the peak hold function.
•
Waveforms shown on the screen and D/A output waveforms are not affected by the peak hold
operation.
• The time when the maximum value occurred is not displayed.
•
The peak hold function does not affect to integrated values.
HIOKI PW8001A961-00
Summary of Contents for PW8001-01
Page 2: ...HIOKI PW8001A961 00...
Page 6: ...Contents iv HIOKI PW8001A961 00...
Page 56: ...50 Checking Connections HIOKI PW8001A961 00...
Page 120: ...114 Power Calculation Method HIOKI PW8001A961 00...
Page 162: ...156 Controlling Integration with External Signals HIOKI PW8001A961 00...
Page 236: ...230 U7005 15MS s Input Unit HIOKI PW8001A961 00...
Page 258: ...252 Index HIOKI PW8001A961 00...
Page 259: ...HIOKI PW8001A961 00...
Page 260: ...HIOKI PW8001A961 00...
Page 261: ...HIOKI PW8001A961 00...
Page 262: ...www hioki com HIOKI PW8001A961 00...