43
Setting Wiring Mode and Configuring Current Sensor Settings
Correcting phase errors for current sensors
Current sensors generally exhibit a tendency for phase error to increase gradually in the high-
frequency region of their frequency band. By using sensor-specific phase characteristics
information to correct the measured values, it is possible to reduce the error component in power
measurement made in a high-frequency region.
Conceptual diagram
Frequency (Hz)
Phase (
°
)
1 M
10
−
30
−
20
−
10
0
10
20
30
100
1 k
10 k
100 k
○: Phase characteristics after correction
Δ
: Phase characteristics before correction
Phase correction for current
sensors with auto-recognition
function
When using a current sensor with the
auto-recognition function, the phase of the
current sensor is automatically corrected.
If you wish to set any phase correction
values, follow the steps listed in the
following “How to enter phase correction
values” section.
How to enter phase correction values
For current sensors without the auto-recognition function, it is recommended to perform phase
correction for the current sensor before measurement.
Display screen
[INPUT]
>
[CHANNEL]
1
1
1
Tap the detailed display area for the
channel to be configured.
2
Tap the
[
Phase Shift
]
area box and
select
[On]
.
When using a current sensor equipped with the
auto-recognition function,
[Auto]
is displayed
as an alternatives. When
[Auto]
is selected,
the correction values are entered automatically.
3
Tap the frequency box, then enter the
frequency with the numeric keypad.
4
Tap the phase difference box, then
enter the phase difference with the
numeric keypad.
5
Tap
[
×
]
to close the setting window.
IMPORTANT
•
Enter the phase correction value accurately.
Invalid settings can cause the correction
process to increase measurement error.
•
Operation outside the frequency range within
which the current sensor’s phase accuracy is
specified is not defined.
2
2 3
3
4
4
5
5
Preparing for Measurement
HIOKI PW8001A961-00
Summary of Contents for PW8001-01
Page 2: ...HIOKI PW8001A961 00...
Page 6: ...Contents iv HIOKI PW8001A961 00...
Page 56: ...50 Checking Connections HIOKI PW8001A961 00...
Page 120: ...114 Power Calculation Method HIOKI PW8001A961 00...
Page 162: ...156 Controlling Integration with External Signals HIOKI PW8001A961 00...
Page 236: ...230 U7005 15MS s Input Unit HIOKI PW8001A961 00...
Page 258: ...252 Index HIOKI PW8001A961 00...
Page 259: ...HIOKI PW8001A961 00...
Page 260: ...HIOKI PW8001A961 00...
Page 261: ...HIOKI PW8001A961 00...
Page 262: ...www hioki com HIOKI PW8001A961 00...