13
1
Overview
1.1 Product Overview
This instrument is a power analyzer that can analyze the power conversion efficiency by
simultaneously measuring the input and output powers of an object under measurement. The
instrument can include up to eight input modules and support various lines to be measured
according to your applications by combining wiring configurations from single-phase to three-phase
four-wire configurations as desired.
1.2 Features
Up to eight modules can be included
By combining two types of input modules with one to eight channels as desired, the measurement system
most suitable for your applications can be configured in a single instrument.
Combination of the two types of input modules enables the configuration of the most
suitable system
Two types of input modules are available: a general-purpose input module with high withstand voltage, the
U7001, and an input module with world-class accuracy of
±
0.03% and world-class high resolution and high-
speed sampling, the U7005.
According to the performance you need, the two types of input modules can be combined and installed on the
PW8001.
Measurement for CAT II at 1500 V is
now possible during the development,
evaluation, and delivery inspection of
power conditioners.
Power measurement basic accuracy
±
0.07%
Efficiency of SiC/GaN inverters and loss
of reactors and transformers can be
measured with high accuracy.
Power measurement basic accuracy
±
0.03% (DC accuracy
±
0.05%)
Sampling frequency
2.5 MHz
15 MHz
ADC resolution
16-bit
18-bit
Measurement frequency band
DC, 0.1 Hz to 1 MHz
DC, 0.1 Hz to 5 MHz
Maximum input voltage
1000 V AC, 1500 V DC
1000 V AC, 1000 V DC
Maximum rated line-to-ground
voltage
600 V AC, 1000 V DC CAT III
1000 V AC, 1500 V DC CAT II
600 V CAT III
1000 V CAT II
Automatically recognizes current sensors
The instrument automatically acquires information on
current sensors connected to itself and corrects their
phase errors.
This significantly reduces the setting time before
measurement and strongly supports accurate power
measurement.
Automatically
acquiring information
Powered by PW8001
Simple Settings (Quick Set)(p. 45)
The Quick Set function allows measurement conditions to be set to typical values at once according to the
selected lines to be measured.
1 Overview
Overview
HIOKI PW8001A961-00
Summary of Contents for PW8001-01
Page 2: ...HIOKI PW8001A961 00...
Page 6: ...Contents iv HIOKI PW8001A961 00...
Page 56: ...50 Checking Connections HIOKI PW8001A961 00...
Page 120: ...114 Power Calculation Method HIOKI PW8001A961 00...
Page 162: ...156 Controlling Integration with External Signals HIOKI PW8001A961 00...
Page 236: ...230 U7005 15MS s Input Unit HIOKI PW8001A961 00...
Page 258: ...252 Index HIOKI PW8001A961 00...
Page 259: ...HIOKI PW8001A961 00...
Page 260: ...HIOKI PW8001A961 00...
Page 261: ...HIOKI PW8001A961 00...
Page 262: ...www hioki com HIOKI PW8001A961 00...