Motor Measurement (Motor Analysis-Equipped Model)
Zero adjustment of motor input
In the following circumstances, perform zero adjustment to eliminate errors caused by input signal
offsets:
•
When an analog DC voltage is inputted to Ch. A, Ch. C, Ch. E, and Ch. G
•
When frequency-based torque signal is input
In the following circumstances, perform zero adjustment while the instrument is receiving zero input
for the torque and RPM signals:
•
When a torque value is displayed even though no torque signal is generated
•
When an RPM value is displayed even though no rotation signal is generated
1
Press the
MEAS
key.
2
Use the
CH
key for channel selection to switch the
display to
[A-D]
or
[E-H]
.
The displayed channel will change each time the
CH
keys are
pressed.
3
Press
0ADJ
.
The confirmation dialog box will be displayed.
4
Tap
[Yes]
.
Zero adjustment is started.
•
When the
[A-D]
channel indicator or
[E-H]
channel indicator is lit up, you can also perform motor
input zero adjustment by pressing the
0ADJ
key even on any page of the Measurement screen.
• Zero adjustment is unavailable for channels with the input setting set to
[Pulse]
.
• Zero adjustment can be performed within an input range of
±
10% of range. Inputting our-of-range
signals cause zero-adjustment to fail.
Displaying Power Numerically
HIOKI PW8001A961-00
Summary of Contents for PW8001-01
Page 2: ...HIOKI PW8001A961 00...
Page 6: ...Contents iv HIOKI PW8001A961 00...
Page 56: ...50 Checking Connections HIOKI PW8001A961 00...
Page 120: ...114 Power Calculation Method HIOKI PW8001A961 00...
Page 162: ...156 Controlling Integration with External Signals HIOKI PW8001A961 00...
Page 236: ...230 U7005 15MS s Input Unit HIOKI PW8001A961 00...
Page 258: ...252 Index HIOKI PW8001A961 00...
Page 259: ...HIOKI PW8001A961 00...
Page 260: ...HIOKI PW8001A961 00...
Page 261: ...HIOKI PW8001A961 00...
Page 262: ...www hioki com HIOKI PW8001A961 00...