118
Factory Default Settings
6.3 Factory Default Settings
The following tables list the instrument’s factory default settings.
The measurement screen and recorded data settings will also be reset.
Parameter
Default setting
Current input
Probe 1
Wiring
1P2W
Synchronization source
U1, U2, U3, U4, U5, U6, U7,
U8 (depends the number of
installed modules)
U range
1500 V
U auto-ranging
Off
U rectifier
RMS
VT ratio
1.0 (off)
I range
Sensor’s rated current
I auto-ranging
Off
I rectifier
RMS
CT ratio
1.0 (off)
LPF
Off
Sensor phase correction
Off
Integration mode
RMS
Upper frequency limit
U7001: 1 MHz
U7005: 2 MHz
Lower frequency limit
10 Hz
ZC HPF
On
Delta conversion
Off
Data refresh interval
50 ms
Measurement mode
Wideband
Grouping
Type 1
THD calculation order
500th
THD calculation method
THD-F
Averaging mode
Off
Power calculation method
Type 1
Efficiency calculation Pin,
Pout
P1
Display language*
Japanese
Beep tone
On
Startup screen selection
Wiring (Wiring screen)
GP-IB address*
1
BNC-and-opt synchronization Off
(Motor) synchronization
source
DC
Motor analysis option wiring
setting
Torque, Speed
Parameter
Default setting
Torque input
Analog
(Motor) LPF
Off
Motor voltage range
5 V
RPM input
Pulse
Torque scaling
1.0
Pulse count
2
Number of motor poles
4
Slip input frequency
fU1
Phase ADJ
0.000
Output range
1 V f.s.
Integration full scale
1
Output items
D/A1 to D/A16: WAVE U1,
I1U2, I2, U3, I3, . . . , U8, I8
D/A17 to D/A20: Trend
Urms1
(depends on the number of
installed modules)
Data saving interval
1 s
Timer
Off
Timer setup
1 min.
Real time control
Off
Text saving format
CSV*
Auto-save
Off
Manual save
Off
Screenshot
Off
Comment entry
Off
Simultaneous saving of
settings
Off
DHCP*
Off
IP address*
192.168.1.1
Subnet mask*
255,255,255.0
Default gateway*
0.0.0.0
RS-232C connection*
RS-232C
RS-232C communications
speed
115200 bps
Time zone
GMT +09:00*
Date format
yyyyMMdd*
Delimiter for date
Hyphen (-)*
HIOKI PW8001A961-00
Summary of Contents for PW8001-01
Page 2: ...HIOKI PW8001A961 00...
Page 6: ...Contents iv HIOKI PW8001A961 00...
Page 56: ...50 Checking Connections HIOKI PW8001A961 00...
Page 120: ...114 Power Calculation Method HIOKI PW8001A961 00...
Page 162: ...156 Controlling Integration with External Signals HIOKI PW8001A961 00...
Page 236: ...230 U7005 15MS s Input Unit HIOKI PW8001A961 00...
Page 258: ...252 Index HIOKI PW8001A961 00...
Page 259: ...HIOKI PW8001A961 00...
Page 260: ...HIOKI PW8001A961 00...
Page 261: ...HIOKI PW8001A961 00...
Page 262: ...www hioki com HIOKI PW8001A961 00...