196
Functional Specifications
Hold function
(1)
Hold
Functions
Stops updating display of all measured values, freezing the presently on-screen figures.
However, updating continues for waveforms, the clock, and on-screen peak-over conditions.
Internal calculations, for example integration and averaging, continue.
Cannot be used in combination with the peak hold function.
Operating mode
Off/on
Action
Pressing the
HOLD
key can activate the function, lighting up the
HOLD
key and displaying the
hold icon on the screen.
Pressing the
PEAK HOLD
key can refresh data when the hold function has been enabled.
Data is refreshed at the internal data intervals (which is distinct from the display refresh
interval).
Output data
Internally retained frozen data is outputted as analog output and stored on a USB flash drive.
(However, waveform output continues).
Backup
None (The function is disabled when the instrument is turned off.)
Constraint
While the hold function is activated, settings that affect measured values cannot be changed.
(2)
Peak hold
Function
The display is refreshed by replacing all measured values with the maximum values obtained
by comparing the absolute values for each measured value (excluding Upk and Ipk). However,
the waveform display and integrated values continue to be refreshed by replacing them with
instantaneous data.
During average operation, the maximum value affects values measured after averaging.
Cannot be used in combination with the hold function.
Operating mode
Off/on
Action
Pressing the
PEAK HOLD
key can activate the function, lighting up the
PEAK HOLD
key and
displaying the peak hold icon on the screen.
Pressing the
PEAK HOLD
key again can turn the function off.
When the hold function is enabled, data is updated when the
HOLD
key is pressed.
Data is refreshed at the internal data intervals (which is distinct from the display refresh
interval).
Output data
During the peak-hold operation, internally retained peak-hold data is outputted as analog
output and stored on a USB flash drive. (However, waveform output continues).
Backup
None (The function is disabled when the instrument is turned off.)
Constraint
While the hold or peak-hold function is activated, settings that affect measured values cannot
be changed.
HIOKI PW8001A961-00
Summary of Contents for PW8001-01
Page 2: ...HIOKI PW8001A961 00...
Page 6: ...Contents iv HIOKI PW8001A961 00...
Page 56: ...50 Checking Connections HIOKI PW8001A961 00...
Page 120: ...114 Power Calculation Method HIOKI PW8001A961 00...
Page 162: ...156 Controlling Integration with External Signals HIOKI PW8001A961 00...
Page 236: ...230 U7005 15MS s Input Unit HIOKI PW8001A961 00...
Page 258: ...252 Index HIOKI PW8001A961 00...
Page 259: ...HIOKI PW8001A961 00...
Page 260: ...HIOKI PW8001A961 00...
Page 261: ...HIOKI PW8001A961 00...
Page 262: ...www hioki com HIOKI PW8001A961 00...