225
U7005 15MS/s Input Unit
10.7
U7005 15MS/s Input Unit
Input specifications
(1)
Specifications common to voltage, current, and power measurement
Sampling
15 MHz, 18-bit
Measurement
frequency band
DC, 0.1 Hz to 5 MHz
Effective
measurement range
1% of range to 110% of range
(2)
Specifications common to voltage measurement
Input terminal profile
Plug-in terminal (safety terminal)
Input method
Isolated input, resistance voltage division
Range
6 V, 15 V, 30 V, 60 V, 150 V, 300 V, 600 V, 1500 V
Crest factor
3 relative to voltage range rating (however, 1.35 for 1500 V range)
Input resistance,
input capacity
4 M
Ω
±
20 k
Ω
, 6 pF typical
Maximum input
voltage
1000 V,
±
2000 V peak
(1300 − f) volts when 400 kHz < f ≤ 1000 kHz (f: frequency of input voltage)
200 V when 1000 kHz < f ≤ 5000 kHz (f: frequency of input voltage)
In the expressions listed above, the unit of
f
is kilohertz.
Maximum rated line-
to-ground voltage
600 V in measurement category III, anticipated transient overvoltage: 6000 V
1000 V in measurement category II, anticipated transient overvoltage: 6000 V
(3)
Specifications common to current measurement
Input terminal profile
Probe 1: Dedicated connector (ME15W)
Input method
Current sensor input method
Range
Probe 1:
40 mA, 80 mA, 200 mA, 400 mA, 800 mA, 2 A
(with a 2 A sensor)
400 mA, 800 mA, 2 A, 4 A, 8 A, 20 A
(with a 20 A sensor)
4 A, 8 A, 20 A, 40 A, 80 A, 200 A
(with a 200 A sensor)
40 A, 80 A, 200 A, 400 A, 800 A, 2 kA
(with a 2000 A sensor)
100 mA, 200 mA, 500 mA, 1 A, 2 A, 5 A
(with a 5 A sensor)
1 A, 2 A, 5 A, 10 A, 20 A, 50 A
(with a 50 A sensor)
10 A, 20 A, 50 A, 100 A, 200 A, 500 A
(with a 500 A sensor)
100 A, 200 A, 500 A, 1 kA, 2 kA, 5 kA
(with a 5000 A sensor)
20 A, 40 A, 100 A, 200 A, 400 A, 1 kA
(with a 1000 A sensor)
Selectable for each wiring
(Only when the same sensors are used for all channels of the same wiring configuration)
Crest factor
3 relative to voltage range rating
Input resistance
1 M
Ω
±
50 k
Ω
Maximum input
voltage
8 V,
±
12 V peak (10 ms or less)
Specifications
HIOKI PW8001A961-00
Summary of Contents for PW8001-01
Page 2: ...HIOKI PW8001A961 00...
Page 6: ...Contents iv HIOKI PW8001A961 00...
Page 56: ...50 Checking Connections HIOKI PW8001A961 00...
Page 120: ...114 Power Calculation Method HIOKI PW8001A961 00...
Page 162: ...156 Controlling Integration with External Signals HIOKI PW8001A961 00...
Page 236: ...230 U7005 15MS s Input Unit HIOKI PW8001A961 00...
Page 258: ...252 Index HIOKI PW8001A961 00...
Page 259: ...HIOKI PW8001A961 00...
Page 260: ...HIOKI PW8001A961 00...
Page 261: ...HIOKI PW8001A961 00...
Page 262: ...www hioki com HIOKI PW8001A961 00...