11.4 Frequently Asked Questions
Why has the instrument saved no measured data even though I measured with the auto-save
setting?
You may have pressed the
RUN/STOP
key.
Press the
START/STOP
key to perform auto-save measurement.
See “Automatically saving measured data” (p. 126).
Why has the message “Unable to automatically generate the filename.” appeared during the auto-
save operation?
Each folder can contain up to 1000 files.
Create another folder to save further files.
See “Recordable time and data” (p. 127).
Why has my computer failed to acquire the MAC address even though I have connected the
instrument to the computer through our LAN.
Check the IP address settings.
Except for the last three digits of the IP address, communications cannot be performed unless all the IP
addresses are set to the same numbers as the computer’s IP address.
See “9.1 Connecting and Setting the LAN Interface” (p. 158).
Can I retrofit channels to my instrument after delivery?
Hioki can modify the instrument by custom order.
Contact your authorized Hioki distributor or reseller.
The saved data includes a numeric value of
1.00E+104
. What does this mean?
This indicates that data points are in a overload condition or a peak-over condition.
Can I use password-protected (secure) USB flash drives with the instrument?
Password-protected USB flash drives cannot be used the instrument.
Use USB flash drives that accommodate the mass storage class standard.
See “7.1 USB Flash Drive” (p. 119).
Why has the instrument failed to detect my USB flash drive?
Cycle the instrument. If the instrument cannot the USB flash drive after the instrument was cycled, try a
different USB flash drive. (Not all USB flash drives are supported.)
See “7.1 USB Flash Drive” (p. 119).
Maintenance and Service
HIOKI PW8001A961-00
Summary of Contents for PW8001-01
Page 2: ...HIOKI PW8001A961 00...
Page 6: ...Contents iv HIOKI PW8001A961 00...
Page 56: ...50 Checking Connections HIOKI PW8001A961 00...
Page 120: ...114 Power Calculation Method HIOKI PW8001A961 00...
Page 162: ...156 Controlling Integration with External Signals HIOKI PW8001A961 00...
Page 236: ...230 U7005 15MS s Input Unit HIOKI PW8001A961 00...
Page 258: ...252 Index HIOKI PW8001A961 00...
Page 259: ...HIOKI PW8001A961 00...
Page 260: ...HIOKI PW8001A961 00...
Page 261: ...HIOKI PW8001A961 00...
Page 262: ...www hioki com HIOKI PW8001A961 00...