79
Motor Measurement (Motor Analysis-Equipped Model)
To connect torque meters and tachometers
Necessary items: L9217 Connection Cord (necessary quantity),
device to be connected (such as torque sensor and tachometer)
1
Verify that the instrument and the device being connected have been turned off.
2
Connect the device’s output terminal to the instrument with a connection cord.
See “Connection examples of motor analysis” (p. 81).
3
Turn on the instrument.
4
Turn on the connected device.
Connection method
There are several different operating modes and connection patterns available for the motor inputs.
Display screen
[INPUT]
>
[MOTOR]
1
1
1
Tap
[Motor analysis option wiring]
to
open the settings window.
2
Select the operation mode for the
motor analysis option channels.
3
Tap
[
×
]
to close the setting window.
2
2
3
3
[Individual input]
mode
The motor inputs can be used as independent analog DC inputs or as pulse inputs.
Operating mode
Settable channels
Description
Individual Input
AB, CD, EF, GH
For measuring
voltage signals and
pulse signals
Using this mode can measure and display the signal from a voltage-output sensor, or measure the frequency
of a pulse input and display the waveform.
Displaying Power Numerically
HIOKI PW8001A961-00
Summary of Contents for PW8001-01
Page 2: ...HIOKI PW8001A961 00...
Page 6: ...Contents iv HIOKI PW8001A961 00...
Page 56: ...50 Checking Connections HIOKI PW8001A961 00...
Page 120: ...114 Power Calculation Method HIOKI PW8001A961 00...
Page 162: ...156 Controlling Integration with External Signals HIOKI PW8001A961 00...
Page 236: ...230 U7005 15MS s Input Unit HIOKI PW8001A961 00...
Page 258: ...252 Index HIOKI PW8001A961 00...
Page 259: ...HIOKI PW8001A961 00...
Page 260: ...HIOKI PW8001A961 00...
Page 261: ...HIOKI PW8001A961 00...
Page 262: ...www hioki com HIOKI PW8001A961 00...