69
Integrating Current and Power
Integration mode
This section describes how to set the integration mode for each channel. The two integration
modes, DC and RMS, are available and can be selected separately for each wiring configuration.
Display screen
[INPUT]
>
[CHANNEL]
1
1
1
To open the settings window, tap the
channel detailed display area.
Detailed settings for each wiring configuration
will be displayed.
2
Tap the
[Mode]
box, then select the
integration mode from the list.
DC
Instantaneous current values and
instantaneous power values for each
sampling are integrated separately
for each polarity.
This can be selected only in 1P2W
wiring mode.
The six items of current integration
(Ih+, Ih−, Ih) and active power
integration (WP+, WP−, WP) are
calculated simultaneously.
RMS
Current RMS value and active
power value acquired at data refresh
intervals are integrated.
Only active power values are
integrated by polarity.
3
Tap
[
×
]
to close the setting window.
2
2
3
3
Displaying Power Numerically
HIOKI PW8001A961-00
Summary of Contents for PW8001-01
Page 2: ...HIOKI PW8001A961 00...
Page 6: ...Contents iv HIOKI PW8001A961 00...
Page 56: ...50 Checking Connections HIOKI PW8001A961 00...
Page 120: ...114 Power Calculation Method HIOKI PW8001A961 00...
Page 162: ...156 Controlling Integration with External Signals HIOKI PW8001A961 00...
Page 236: ...230 U7005 15MS s Input Unit HIOKI PW8001A961 00...
Page 258: ...252 Index HIOKI PW8001A961 00...
Page 259: ...HIOKI PW8001A961 00...
Page 260: ...HIOKI PW8001A961 00...
Page 261: ...HIOKI PW8001A961 00...
Page 262: ...www hioki com HIOKI PW8001A961 00...