152
Waveform/Analog Output (Waveform and D/A Output Option)
Output retained
Output retained
Time
Output retained
(3)
(4)
(1)
+5 V
0 V
−5 V
(2)
Hold
Integration
reset
Integration
stop
Integration
stop
Integration
start
Integration
restart
Integration
restart
Hold
canceled
Hold
Integration
start
Integration
stop
(1) Analog output varies with the start of integration. Analog output is retained when integration
stops.
(2) If the voltage for the integrated value exceeds
±
5 V, analog output will switch to 0 V before
continuing to vary again.
(3) Analog output is retained when the display is frozen during integration. When the hold
operation is canceled, analog output will vary based on the original integrated value.
(4) When the integrated value is reset, analog output will switch to 0 V.
HIOKI PW8001A961-00
Summary of Contents for PW8001-01
Page 2: ...HIOKI PW8001A961 00...
Page 6: ...Contents iv HIOKI PW8001A961 00...
Page 56: ...50 Checking Connections HIOKI PW8001A961 00...
Page 120: ...114 Power Calculation Method HIOKI PW8001A961 00...
Page 162: ...156 Controlling Integration with External Signals HIOKI PW8001A961 00...
Page 236: ...230 U7005 15MS s Input Unit HIOKI PW8001A961 00...
Page 258: ...252 Index HIOKI PW8001A961 00...
Page 259: ...HIOKI PW8001A961 00...
Page 260: ...HIOKI PW8001A961 00...
Page 261: ...HIOKI PW8001A961 00...
Page 262: ...www hioki com HIOKI PW8001A961 00...