184
Specifications of Input, Output, and Measurement
(4)
Frequency measurement specifications
Number of
measurement
channels
Up to 8 channels (fU1 to fU8, fI1 to fI8), depending on the number of installed modules
Measurement
method
Reciprocal method
The waveforms processed with the zero-cross filter are measured.
Measurable range
0.1 Hz to 2 MHz (The display shows
0.00000 Hz
or
– – – – – – Hz
if measurement was not
possible.)
The range is limited by the input module’s measurement band and the measurement lower
frequency limit setting.
Measurement
accuracy
±
0.005 Hz
Assuming all the following conditions are met:
•Measurement parameter: voltage frequency
•Data refresh interval:
50 ms or more
•Voltage range:
15 V range or higher
•Inputted waveform:
a sine wave with a magnitude of at least 50% of the range
•Frequency range:
45 Hz to 66 Hz
Under conditions other than listed above:
±
0.05% of reading
(With a sine wave at least 30% of the measurement range of the measurement source)
Display resolution
0.10000 Hz to 9.99999 Hz, 9.9000 Hz to 99.9999 Hz,
99.000 Hz to 999.999 Hz, 0.99000 kHz to 9.99999 kHz,
9.9000 kHz to 99.9999 kHz, 99.000 kHz to 999.999 kHz,
0.99000 MHz to 2.00000 MHz
(5)
Integration measurement specifications
Measurement mode
Can be chosen between RMS and DC for each wiring.
(The DC mode is selectable for the 1P2W wiring configuration only.)
Measurement item
Current integration (Ih+, Ih−, Ih), Active power integration (WP+, WP−, WP)
The instrument measures Ih+ and Ih− only in DC mode; Ih only in RMS mode.
Measurement
method
Digital calculation based on current and active power. (Calculations are performed using
not-averaged values during averaging.)
In DC mode: Integrates current values and instantaneous power values for each polarity at
every sampling point.
In RMS mode: Integrates current RMS values and active power values at the
measurement intervals.
Only the active power is integrated by polarity. (Active power is integrated by polarity at
every period of the synchronization source.)
(The sum of integrated effective power values of a polyphase wiring configuration is the
sum of effective power values by polarity at the measurement intervals.)
Measurement
interval
Same as the data refresh interval
Display resolution
999999 (6 decimal point),
Starts from the resolution assuming 1% of each range to be 100% of range.
Measurable range
0 to
±
99.9999 PAh
0 to
±
99.9999 PWh
Integration time
0 s to 9999 h 59 min. 59 s
Integration stops if the integration time exceeds the range.
Integration time
accuracy
±
0.02% of reading (−10
°
C to 40
°
C)
Integration accuracy
±
(Current or active power)
±
(Integration time accuracy)
Backup function
None
If a power outage occurs during integration, the integration stops after power is restored
and the integration data is reset.
HIOKI PW8001A961-00
Summary of Contents for PW8001-01
Page 2: ...HIOKI PW8001A961 00...
Page 6: ...Contents iv HIOKI PW8001A961 00...
Page 56: ...50 Checking Connections HIOKI PW8001A961 00...
Page 120: ...114 Power Calculation Method HIOKI PW8001A961 00...
Page 162: ...156 Controlling Integration with External Signals HIOKI PW8001A961 00...
Page 236: ...230 U7005 15MS s Input Unit HIOKI PW8001A961 00...
Page 258: ...252 Index HIOKI PW8001A961 00...
Page 259: ...HIOKI PW8001A961 00...
Page 260: ...HIOKI PW8001A961 00...
Page 261: ...HIOKI PW8001A961 00...
Page 262: ...www hioki com HIOKI PW8001A961 00...