6.2 Switching the Unit On and Off
46
S
W I T C H I N G O N
Operation
Activate Charge/ Discharge
Close contactors(Relais-Button)
Activate channel(On-Off Button)
Set values for V, I and P,
activate pause, choose control
mode
check the max. set-values and
and monitored values
Selecting the desired mode of
operation(Batt-Sim/Batt-Test)
Always set all set-values for V, I and P. I and P have to be nega-
tive, when the system has to take in energy, even when “Dis-
charge” or “Sink” has been chosen as mode of operation. The
set-values for the uncontrolled variables should be set gener-
ously so that the control has sufficient freedom.
Set the set-value and monitoring limits so that your test object
is not endangered
When activating a channel both AFEs are first precharged and
then started.
Current is activated