ServiceMode
WTX120
A4500-1.1
HBM: public
59
→ Calibrate
AOut 1: 0V =
9
Calibrate zero point, e.g. =0V, adjust analog sig
nal in increments
AOut 1: 10V =
4095
Calibrate full load value, e.g. =10V, adjust analog
signal in increments
Information
The changed values are overwritten when the mode is changed!
Information
After resetting to factory defaults, the configuration for the
Analog outputs
menu must be performed.