Calibration
WTX120
A4500-1.1
HBM: public
129
Menu
Key
Function/Secondary selection
Select Group 1-9
Press twice for group
2 Calibration
→ Linearization points present
Fixpoints active!
Linearization entries have already been made;
they may influence the calibration.
Menu
Key
Function/Secondary selection
Geo Value
99
Enter the geo value. Values between 0 and 31
can be set, according to the geo value table. The
geo value of the
calibration
location is set. If dif
fering, the geo value of the
installation
location
can be entered subsequently.
Calibrate Zero?
Y
Zero calibration of the scale: Relieve load on
scale
Y
Calibrate zero point
N
Continue to
Zero(mV/V)
step
Calibrating...
Measure the zero point of the scale
Zero:
0.00
Display the 10 times higher weight value resolu
tion (as a check)
Zero(mV/V):
0.40000
Display the standardized signal - see
Calculate Span
.
Hinweis:
The signal can be calculated in millivolts
from the displayed mV/V value by multiplying it by
the
Excitation
value (5V or 10V).
Information
The Clr key can be used to clear the value and enter a new one.
Summary of Contents for WTX120
Page 1: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 3: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 184: ...Service password 182 A4500 1 1 HBM public WTX120 ...
Page 185: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 366: ...Service Passwort 182 A4500 1 1 HBM public WTX120 ...
Page 367: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 549: ...Mot de passe de service WTX120 A4500 1 1 HBM public 183 ...