![HBM WTX120 Operating Manual Download Page 130](http://html1.mh-extra.com/html/hbm/wtx120/wtx120_operating-manual_2111754130.webp)
Calibration
128
A4500-1.1
HBM: public
WTX120
11.5
Calibration
For more information on the keys necessary to access this menu group, go to
the
.
With this menu group the scale is calibrated. In addition to calibration at full
load, part load calibration is also possible. If you want to move the scale to a
different location after calibrating it, you can use the geo value to compensate
for the differing gravitational pull, without having to recalibrate the scale (not for
applications subject to weights and measures legislation). Precalibration with
out weights is also possible by entering the standardized signal (mV/V).
Example of calibration of a tank weigher:
S
Three 1000kg load cells with sensitivity of 1.99995mV/V, 2.00005mV/V and
2.00000mV/V respectively are used.
S
The maximum load is 1500kg; the interval 0.5kg
S
The weight of the empty tank is 600kg.
Zero point
Load point
Fig. 12.4
Example of calibration of a tank weigher
Summary of Contents for WTX120
Page 1: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 3: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 184: ...Service password 182 A4500 1 1 HBM public WTX120 ...
Page 185: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 366: ...Service Passwort 182 A4500 1 1 HBM public WTX120 ...
Page 367: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 549: ...Mot de passe de service WTX120 A4500 1 1 HBM public 183 ...