Calibration
138
A4500-1.1
HBM: public
WTX120
Function/Secondary selection
Key
Menu
PowerUp Zero +/-
10%
Power-up zero setting range
This sets the range around the calibrated zero
point in which the scale is automatically zeroed
on power-up.
OFF
Power-up zero off
±2%, ±10%
Range size
Overload
9999
Here the overload display suppression (in D) is
set. If the scale exceeds the value (maximum
load + overload), the display is hidden.
Example:
Overload 9
On multiple-range scales, 1d corresponds to the
digits of the lowest weighing range.
NTEP approval?
N
Switch to the metrological properties according to
American requirements (National Type Evalua
tion Program) as per Handbook 44.
N
Off (default)
Y
On
Must be set to
Y
(es) for applications in the USA,
Canada and Australia!
Underload 20d?
Y
Hide weight display if gross weight values less
than -20d (below zero)
N
Off
Y
On
For applications in accordance with OIML R76,
Underload 20d
must be set to
Y
Summary of Contents for WTX120
Page 1: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 3: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 184: ...Service password 182 A4500 1 1 HBM public WTX120 ...
Page 185: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 366: ...Service Passwort 182 A4500 1 1 HBM public WTX120 ...
Page 367: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 549: ...Mot de passe de service WTX120 A4500 1 1 HBM public 183 ...