Calibration
122
A4500-1.1
HBM: public
WTX120
S1
Switch up:
calibration data saved, no changes possible
Switch down:
(direction of terminals)
calibration data can be changed
S1
WTX
120
Fig. 12.3
Calibration switch S1 WTX120
Information
Position of switch S1 (under covering)
For applications subject to weights and measures legislation, the covering and
lid must be sealed by adhesive labels so that the position of the calibration
switch can no longer be changed.
→ If switch S1 up / saved
Calibration Locked
Switch S1 not in calibration position, parameters
cannot be saved!
Access calibration without save option (e.g. for
checking)