Calibration
WTX120
A4500-1.1
HBM: public
137
Information
The filtration level can also be changed retrospectively in
ServiceMode
and
Mastermode
. This makes it possible to change the filtration level even after
legal verification, without breaking the verification seal.
The filter setting in
ServiceMode
overwrites the calibration value every time
the application is started. So to check the validity of the setting the value in
ServiceMode
must be checked!.
Menu
Key
Function/Secondary selection
Auto Zero Range
0.5D
Automatic zeroing
This sets the range around the zero point in
which the weight value must be in order for auto
matic zeroing to activate.
OFF
Turn off zeroing
0.5D, 1.0D, 3.0D
Range size
PbZero (%) +
999
Zero setting range (+)
This sets the range above the power-up zero
point in which the scale can be zeroed by the
Zero key, and in which automatic zeroing takes
effect.
PbZero (%) -
999
Zero setting range (–)
This sets the range below the power-up zero
point in which the scale can be zeroed by the
Zero key, and in which automatic zeroing takes
effect.
Summary of Contents for WTX120
Page 1: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 3: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 184: ...Service password 182 A4500 1 1 HBM public WTX120 ...
Page 185: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 366: ...Service Passwort 182 A4500 1 1 HBM public WTX120 ...
Page 367: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 549: ...Mot de passe de service WTX120 A4500 1 1 HBM public 183 ...