Calibration
132
A4500-1.1
HBM: public
WTX120
11.6
Linearization
For more information on the keys necessary to access this menu group, go to
the
.
Linearization compensates for any variation of the scale characteristic (from
the ideal straight line) by setting up to 6 correction points adapted to the actual
scale characteristic.
The linearization software supports up to 6 correction points, though in practice
1 or 2 are almost always enough. The linearization points should be placed at
the points on the weight curve where the greatest variations occur. Lineariza
tion points must be lower than the set maximum load. The lines between adja
cent linearization points are calculated as straight, meaning the variations are
interpolated in linear mode.
Menu
Key
Function/Secondary selection
Select Group 1-9
Press 3 times for group
3 Linearization
→ Linearization points already present
Fixpoint 1:
999999
Display existing linearization points (1 to max. 6)
Delete existing linearization point
Calibrate Fixpt?
N
Lin.Signal
0.50000
Linearization signal is displayed
Summary of Contents for WTX120
Page 1: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 3: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 184: ...Service password 182 A4500 1 1 HBM public WTX120 ...
Page 185: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 366: ...Service Passwort 182 A4500 1 1 HBM public WTX120 ...
Page 367: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 549: ...Mot de passe de service WTX120 A4500 1 1 HBM public 183 ...